Axetris AG
LGD Compact Integration Guide V1.8
04.12.2019
23/34
Without housing
Figure 17
Ground concept without housing
Notes:
The cable shield from the sensor module has to be connected to the earth potential.
) have to be connected to the same earth potential.
Without housing and without interface board
Figure 18
Ground concept without housing and without interface board
Notes:
The cable shield from the sensor module has to be connected to the earth potential.
) have to be connected to the same earth potential.