Install the test set-up as shown:
Set the PIM-P to Field Mode
Set frequencies of carrier signal 1 and 2 to the start and end frequency of the
provided frequency band.
Set the power level of each signal to 20 dBm (100mW) – with ALC: ON
Set the Filter bandwidth of the PIM-P to 1.2 kHz (default).
To avoid damage, RF Power of the PIM-P must be OFF
Set Signal Generator to IM3 frequency.
Apply signal to PIM-P (-60dBm to -120dBm)
RSSI should measure the applied power
If the reading is off by more than +/- 0.8 dB but not worse than +/-15 dB,
it may be an indication that receiver is out of calibration. If the
measurement shows a level lower than -125dBm, the receiver does not
detect the frequency at all. It may be asynchronous. In the cases described
above, contact your nearest AWT Service Center.
With the reading in limits, continue procedure:
• Change the frequency of the RF Generator by 2.5 kHz.
• The receiver will filter the signal out and the RSSI reading of PIM-P should be
lower than -125dBm.
With the reading in limits, continue procedure:
• Increase the PIM-P Signal frequency F1 by 1 MHz
• Set Signal Generator to the frequency that correlates to the IM3 product of
frequencies F1 and F2
• Read the RSSI on PIM-P; the value should be equal to +/-1 dB of Signal Generator’s
Power level.
• Repeat the process you are no longer able to increase the signal frequency F1 due to
band restrictions
• If done, set F1 back to start frequency
• Decrease PIM-P Signal frequency F2 by 1 MHz
• Set Signal Generator to Frequency that results for IM3 product of frequencies F1
and F2
Содержание PIM-P
Страница 4: ...1 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved INTRODUCTION 1...
Страница 6: ...2 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved SAFETY SUMMARY 2...
Страница 10: ...3 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved PIM P ELEMENTS 3...
Страница 13: ...4 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved PRECAUTIONS 4...
Страница 20: ...5 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved POWERING PIM P UP DOWN 5...
Страница 23: ...6 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved GETTING STARTED 6...
Страница 27: ...7 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved PIM P OPERATION 7...
Страница 78: ...9 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Pull Down Menu 9...
Страница 85: ...10 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved What is PIM 10...
Страница 88: ...11 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved PIM P Technical Information 11...
Страница 92: ...12 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Maintenance 12...
Страница 93: ...12 2 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved 12 1 Block Diagram...
Страница 100: ...13 1 PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Accessories 13...
Страница 102: ...i PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Appendix...
Страница 105: ...iv PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Appendix B EC Declaration of Conformity...
Страница 106: ...v PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved NRTL Declaration of Conformity...
Страница 107: ...vi PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Appendix C Lid Hold...
Страница 108: ...vii PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Lid Release...
Страница 109: ...viii PIM P Analyzer User s Manual Copyright AWT Global LLC All rights reserved Foot...