46
ENTER. The “CLEAN SWEEP” screen will appear. This function can take up
to 45 seconds to complete – DO NOT TURN OFF OR INTERRUPT POWER
DURING THIS OPERATION!
WARNING: Performing a “CLEAN SWEEP” operation will erase ALL
programmed information including K-factors, linearizer tables and output
limits programming. Current loop calibrations will also be set to default
values and will need to be reset. Use this function ONLY when all other
troubleshooting methods have failed and unstable operation of the controller
remains. It is strongly suggested that the status screen be viewed prior to
performing a restore defaults to see if the raw data shown on it looks nominal
to the current process. Also make sure the K-Factor or linearizer table data is
available for reprogramming as these values will all be lost.