Figure 15: Total Source
Zero Total mA
Enter the mA value to represent a zero totalizer value. Minimum allowable value
is 2mA.
Max Total Value
Enter the maximum totalizer value to monitor.
Max Total mA
Enter the mA to represent the above entered maximum totalizer value. 20mA is
the maximum allowable value.
GRAND TOTAL source
Figure 16: Grand Total Source
GRAND TOTAL source allows the user to configure the mA output to represent
the Grand Total value. If the Grand Total value goes above the programmed
maximum Grand Total value, the analog output will continue to increase, up to a
maximum of 22mA. This is useful for system fault detection.
Zero Grand Total mA
Enter the mA value to represent a zero Grand Total value. Minimum allowable
value is 2mA.
Max Grand Total Value
Enter the maximum Grand Total value to monitor.