One end of the 50 Ohm resistance is connected to ground, and access to the other end
is provided through a coaxial connector. This output should be terminated with 50
Ohms, and connected to a high-bandwidth (> 400 MHz) oscilloscope. The voltage
across this resistance is directly proportional to the current through the diode. By
observing the current waveform through the diode, the reverse recovery time may be
determined.
While the provided test jig is intended to be flexible and easy to use, users can also
develop their own test jigs easily.
The AVR-EB4-B includes an internal trigger source, but it can also be triggered or
gated by an external source. A front-panel pushbutton can also be used to trigger the
instrument.
The AVR-EB4-B features front panel keyboard and adjust knob control of the output
pulse parameters along with a four line by 40-character backlit LCD display of the
output amplitude, pulse repetition frequency, and delay. The instrument includes
memory to store up to four complete instrument setups. The operator may use the front
panel or the computer interface to store a complete “snapshot” of all key instrument
settings, and recall this setup at a later time.
This instrument is intended for use in research, development, test and calibration
laboratories by qualified personnel.
7
Содержание AVR-EB4-B
Страница 45: ...The next photo shows a MELF device installed between the two pins 45 A MELF packaged Device Under Test DUT...
Страница 57: ...Then open the tan colored DUT socket by flipping forward the two black latches on the front edge of it 57...
Страница 69: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Страница 71: ...PCB 104G KEYPAD DISPLAY BOARD...
Страница 76: ...TEST JIG WIRING AVX TRR HPOST...
Страница 83: ...PERFORMANCE CHECK SHEET 83...