Ohm terminator is suggested for this purpose.)
6. GPIB Connector
. A standard GPIB cable can be attached to this connector to allow
the instrument to be computer-controlled. See the “Programming Manual for -B
Instruments” for more details on GPIB control.
7. RS-232 Connector.
A standard serial cable with a 25-pin male connector can be
attached to this connector to allow the instrument to be computer-controlled. A user
name (“admin”) and a password (“default”, as shipped from the factory) are required
when logging into a serial terminal session. The internal controller attempts to auto-
sense the parity setting. It may be necessary to send a few return characters before
attempting a login in order to provide enough data to allow this auto-sensing to work.
(A standard Linux “agetty” process is used to implement serial control internally.) See
the “Programming Manual for -B Instruments” for more details on RS-232 control.
8. Network Connector
. This Ethernet connector allows the instrument to be remotely
controlled using the VXI-11.3, ssh (secure shell), telnet, and http (web) protocols. See
the “Programming Manual for -B Instruments” for more details.
9. PULSE OUT CONNECTOR
. This BNC connector provides the pulse output signal to
the test jig. This output should be connected to the corresponding input on the test jig
using one of the the supplied coaxial cables (60 cm or 5 m).
Caution: Voltages as high as 240V may be present on the center conductor of this
output connector. Avoid touching this conductor. Connect to this connector using
standard coaxial cable, to ensure that the center conductor is not exposed.
10.CONTROL Connector
. This DB-9 female connector should be connected to the
corresponding connector on the test jig using the supplied DB-9 cable. This cable
contains the safety interlock signals that ensure that the test jig lid is closed. The
pinout is as follows:
Pin 1 - To test jig switch 1.
Pin 2 - To test jig switch 2.
Pin 5 - Ground.
Pin 6 - To test jig switch 1.
Pin 7 - To test jig switch 2.
Pin 9 - Safety sensor power supply (+15V through 680 Ohms).
When the test jig lid is safely closed, Pin 1 is shorted to Pin 6, and Pin 2 is shorted to
Pin 7.
20
Содержание AVR-EB4-B
Страница 45: ...The next photo shows a MELF device installed between the two pins 45 A MELF packaged Device Under Test DUT...
Страница 57: ...Then open the tan colored DUT socket by flipping forward the two black latches on the front edge of it 57...
Страница 69: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Страница 71: ...PCB 104G KEYPAD DISPLAY BOARD...
Страница 76: ...TEST JIG WIRING AVX TRR HPOST...
Страница 83: ...PERFORMANCE CHECK SHEET 83...