![Aventech AIMMS-20 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/aventech/aimms-20/aimms-20_operating-manual_3028858037.webp)
Rev:
1.00
Date:
11.22.2013
Page:
37 of 40
AIMMS-20
OPERATING MANUAL
Document-P/N:
AIMMS20-OM
© Aventech Research Inc. 2013. All Rights Reserved.
A.3.3 Data Block, Purge Flow Packet (ID = 4)
Byte #
Description
Var. Type
Scale
5
Flow rate, mL/min – lsb
Int
1
6
Flow rate, mL/min – msb
Int
1
A.3.4 Data Block, Internal Probe Temperature Packet (ID = 5)
Byte #
Description
Var. Type
Scale
5
Heater block temp. [C], fwd – lsb
Int
10
6
Heater block temp.[C], fwd – msb
Int
10
7
Heater block temp. [C], aft – lsb
Int
10
8
Heater block temp. [C], aft – msb
Int
10
9
Low-temp. threshold [C], lsb
Int
10
10
Low-temp. threshold [C], msb
int
10
Note:
Low temperature threshold setting is the temperature below which the internal
probe block heaters are engaged. The high threshold, i.e. temperature above which
heaters are turned off, is always set to the low thr 10 Celsius.