ID
:
020392
Revision:
1-01
5
© Avantes BV
SMA-plug
Connect the SMA connector of the fiber to the SMA-plug of the Light Source.
I/O connector
For automatic disable control and long-life or high-power to be controlled by software it is needed
to connect an AvaSpec spectrometer to the AvaLight-HAL-Mini with the interface cable IC-DB26-2.
Now it is possible to control the output operation mode by software (AvaSoft).
The signals on the I/O connector are described below.
Pin configuration
Pin Name
Description
5
High
power
Input
Input, used to increase the power output with 10% (reduced life
time of the bulb)
8
Vout
Output
Nominal 12VDC output (minimum 10.8V, maximum 13.2V)
10
GND
Ground
11
Disable
Input
Input, used to disable/shut down the output
12 Long-
life
Input
Input, used to decrease the power output with 10% (increased life
time of the bulb)
Below the electrical characteristics for the input signals.
Electrical characteristics
Name
Low level detection High level detection Maximum voltage
Disable
< 1.0V
> 2.0V
13.2V
Long life
< 1.0V
> 2.0V
13.2V
High power
< 1.0V
> 2.0V
13.2V
Note:
If high-power and long-life are both high, the output power is not specified
2.2
12VDC Power plug
The power supply specifications are mentioned in the table below
Name
Minimum
Typical
Maximum
Voltage
10V
12V
15V
Power
18Watt