16
STEP 9 b: Calibration of the Redox sensor
- Remove the sensor from its soaking water indicated in Section 4.3, Step 2, and remove any excess water.
PLACE THE SENSOR IN THE 475 mV SOLUTION, SHAKE, LEAVE INSIDE WITHOUT TOUCHING
pH 8.5 PRESS OK
Hold down the
CAL/OK
button for 5 SECONDS.
CALIBRATION mV 475
462 pH (-90s)
QUALITY 100%
475 pH PRESS OK
Tip
PRESS THE ESC KEY AT ANY
TIME TO EXIT THE
CALIBRATION PROCEDURE
Tip
PRESS THE ESC KEY AT ANY
TIME TO EXIT THE
CALIBRATION PROCEDURE
After pressing
OK,
the unit
continues operating normally