If the CPU is busy with other programs, printing, or lots of traffic on 

the USB bus, then signals to and from the ValveLink8.2 may suffer.  For 

maximum performance, plug the ValveLink8.2 into a USB port connected 

to its own USB bus.  Remember that computers with multiple USB ports 

may simply be splitting a single USB bus internally.  Check with your 

computer's manufacturer or add a separate USB card to your computer for 

minimum latency.

The System Status window has a large switch labeled "Local Control" 

and "Remote Control."  When set to "Local," each ValveLink8.2 follows 

its own Mode settings as 

described in the "Modes" 

chapter earlier in this 

manual.  When set to 

"Remote Control," all 

networked ValveLink8.2s 

behave according to a single 

common "Mode" setting 

made on the System Status 

window.  In this way, up 

to eight ValveLink8.2s can 

be networked into a single, 

virtual 64-channel controller.  

A single "Master Valve" can 

be selected on a "Master 

Controller" like Mode #5 on the ValveLink8.2.  Also "1-on" Mode means 

only one valve will be open at-a-time on all of the networked controllers.

When the System Status switch is set to "Local Control" you can click the 

"Mode" button on any ValveLink8.2 sub-window to access a dialog box 

of Modes for that ValveLink8.2 along with eight text "labels" which will 

display under each button.  These labels can be used to help identify which 

solutions are being delivered through each valve.

The same comments about USB communications speed above also apply 

to multiple-ValveLink8.2 networking.  If "1-on" Mode is used and a button 

is pressed on one ValveLink8.2, that signal must travel to the User Interface 

program on the PC and back to all other ValveLink8.2s to close any 

open valves.  The double delay may be noticeable on slower computers 

or USB busses.

AutoMate Scientific chose the older USB-1 protocol over the newer, high-

speed USB-2 architecture for three reasons:  1) less expensive, 2) older 

computers don't have USB-2 ports, and 3) USB-2 has more bandwidth 

(transfers large amounts of data faster), but the same latency (delay) as 

USB-1.  The ValveLink8.2 needs short bursts of data quickly, so USB-2 

would not help.

*This software requires Windows .NET version 4.0 or higher be 

installed.

31

30

Содержание ValveLink 8.2

Страница 1: ...ValveLink8 2 User s Manual AutoMate Scientific R E A D Y F O R R E S E A R C H TM...

Страница 2: ...ve Tubing Valve Care Perfusion Pencil Flow Rates Perfusion Pencil Multi barrel Manifold Perfusion Pressure Kit Josef Kewekordes II Heath Lukatch Dave Blau ValveLink8 2 Part 01 18 For indoor use only f...

Страница 3: ...RCA jacks on the back of the box Why not connect valves directly to your computer Several reasons 1 Most computer interfaces provide 5V signals not strong enough to drive most valves 6 to 12V DC and h...

Страница 4: ...wheels and pipette pullers ValveLink8 2 Laboratory Applications Perfusion A ValveLink8 2 can be programmed to precisely switch solutions perfusing over biological samples for physiological research T...

Страница 5: ...on inline heater plus multi channel focal drug delivery maintain bath temperature and rapid drug wash out with a high flow bath line while quickly switching 4 8 preheated solutions through the Perfusi...

Страница 6: ...ld also see the photo on page 5 Find additional photos and diagrams of the pressure system on page 15 2 3 4 1 Assemble ringstand base and rod 2 Attach valve unit 3 Insert syringes with stopcocks Pinch...

Страница 7: ...ine directly to the outflow of the manifold for running cleaning solutions and drying your lines quickly Determine dead volume from the manifold to prep by measuring time and volume required to flush...

Страница 8: ...e Tips reservoir height 40cm Gravity 0 2psi air pressure 0 07ml min 4 6ml hour 8psi air pressure 0 25ml min 15ml hour PTFE Micro manifold Use To use the micro manifold you either need pinch valves or...

Страница 9: ...and removable tips are shipped non sterile They can be chemically sterilized or autoclaved However repeated autoclaving may weaken the adhesive bond between the luer lock fitting and the needle Maint...

Страница 10: ...es mirroring the computer signals low off high on with one millisecond accuracy Usually data acquisition software like pClamp Pulse LabView etc or custom user software is programmed to send digital si...

Страница 11: ...nt panel analog output BNC as an event marker every time a valve opens or closes The marker voltage occurs regardless of how the valve was opened or closed i e manual pushbutton or computer digital in...

Страница 12: ...an electrical path for the spill sensor circuitry In this way a drop of liquid passing the sensor strip appears as a sharp drop in electrical resistance between the two otherwise disconnected copper c...

Страница 13: ...traces A similar conductive epoxy is also available It makes the connection of two wires to your silver traces even easier Simply draw two parallel silver lines approximately 1 16 1 mm apart on any s...

Страница 14: ...between two ValveLink8 2s and control up to 16 valves with a single 4 bit TTL source Set the first ValveLink8 2 to Mode 2 set the second to Mode 3 and connect TTL input bits 1 4 from both ValveLink8 2...

Страница 15: ...ur version of Windows the first time a ValveLink8 2 is installed by USB cable Windows will find new hardware A dialog box will appear asking if you want Windows to connect to Windows Update to search...

Страница 16: ...on rotary switch on the back of each ValveLink8 2 next to its USB port If you connect multiple ValveLink8 2s to the computer be sure to set each one to a different address If you connect a ValveLink8...

Страница 17: ...one valve will be open at a time on all of the networked controllers When the System Status switch is set to Local Control you can click the Mode button on any ValveLink8 2 sub window to access a dia...

Страница 18: ...ar as BNC connectors on the front panel Digital outputs 4 7 are located on the back panel of newer DigiData interfaces Older DigiData 1200 s only have four outputs 2 The Modes chapter explains the two...

Страница 19: ...tubing at an angle and wetting it makes it easier to insert Backflow When a manifold is used in a plumbing arrangement none of its ports can be left unconnected or liquid will simply flow backwards o...

Страница 20: ...g see 1 above and the section below If it qualifies for all three of the above AutoMate Scientific will accept the valve for subsequent return to the factory Repair or replacement is at their discreti...

Страница 21: ...e Safety and Handling data sheets for the above described chemicals are enclosed c We accept full responsibility for any injury to AutoMate Scientific employees or employees of the valves original man...

Страница 22: ...e power requirements of this product 6 Do not allow anything to rest on the power cord Do not locate this product where people will walk on the cord 7 If you have to use an extension cord with this pr...

Страница 23: ...nsor licensees be liable for contingent special direct indirect or consequential damages for the breach of any express or implied warranty or resulting from the use failure or malfunction of any produ...

Страница 24: ...entific Inc 812 Page St Berkeley CA 94710 U S A 800 998 MATE 6283 international 510 845 6283 fax 510 280 3795 e mail info autom8 com www http www autom8 com AutoMate Scientific R E A D Y F O R R E S E...

Отзывы: