background image

Pinch Valve Tubing

Pinch valves have no fittings. Simply slide larger tubing over small pinch 

valve tube above and below the valves.  The tubing in each pinch valve 

can be replaced for easy cleaning.  It is 

1

/

32

" i.d. x 

1

/

16

" o.d. silicone tubing.  

We have only found one type of tubing this size which works well in these 

pinch valves.  AutoMate Scientific sells this replacement tubing as part 

#05-14.  The tubing can be removed from the valves even when they are 

closed.  Remember that any liquid remaining in the tube may spill when it 

is removed from the valve.  New tubing can be installed while the valves 

are closed, but it is much easier if they are opened/energized.  If you are 

concerned about the same piece of tubing being squeezed for long periods 

of time, you can slide the silicone tubing up or down in the valve by a 

few millimeters to expose a new section to the plunger.  The same piece of 

tubing can be used for years without fatigue.

Valve Care

Rinse PTFE valves every day.  Do not get pinch valves wet.  See sections 

on Faulty Valves, Stuck Pinch Valves, Valve Cables, Replacing Valves and 

Leaks in the "Techniques" chapter later in this manual.

Perfusion Pencil Flow Rates

360µm Removable Tips  (reservoir height 40cm)

Gravity (full reservoir)             1.15ml/min  +/- .04

2psi air pressure                  

3.98ml/min  +/- .08

8psi air pressure                 

10.66ml/min  +/- .27

250µm Removable Tips  (reservoir height 40cm)

Gravity (empty reservoir)            0.35ml/min  +/- .05

Gravity (full reservoir)             0.45ml/min  +/- .05

2psi air pressure                 

1.66ml/min  +/- .14

8psi air pressure                 

5.16ml/min  +/- .04

100µm Removable Tips (reservoir height 40cm)

Gravity                                          0

2psi air pressure                   

0.07ml/min = 4.6ml/hour

8psi air pressure                    

0.25ml/min = 15ml/hour

PTFE Micro-manifold Use

To use the micro-manifold, you either need pinch valves or PTFE valves 

equipped with 

1

/

16

" o.d. outflow hose barbs.  Micro-manifolds are shipped 

with a short piece of 

1

/

16

" o.d. tubing in each hole (inputs and outflow).  

You can either connect the valves to manifold inports with a piece of 

1

/

16

i.d. tubing over both the valve barbs (or pinch tubing) and short pieces of 

manifold tubing:

Valve with 

1

/

16

" o.d. barb or pinch tubing

v

1

/

16

" i.d. tubing over barb/pinch tube

and over manifold short tubing

v

Manifold with short pieces of 

1

/

16

" o.d. tubing in each port

... or, for even less dead volume, use a short piece of 

1

/

16

" i.d. tubing over 

the valve barb (or pinch tubing), and piece of smaller 

1

/

16

" o.d. tubing 

inserted inside the secondary tube and manifold ports (you may need to 

supply extra small PE-160 or PTFE tubing for this option):

Valve with 

1

/

16

" o.d. barb or pinch tubing

v

Short piece of 

1

/

16

" i.d. tubing over valve barb/pinch

v

Small diameter 

1

/

16

" o.d. tubing inserted inside valve tube

and inserted inside manifold ports

v

Manifold without connector tubing.  

1

/

16

" holes.

Cutting the tubing at an angle and wetting it will make insertion easier.  

Cutting the tubing square, however, will minimize dead volume when 

inserted as far as possible into the manifold – being careful not block at the 

point of convergence.  Remember to keep all tubing as short as possible.  

Use the provided Allen wrench and small screw near the manifold outflow 

to adjust flow rate.  AutoMate Scientific now sells small PTFE manifold 

'stoppers' for plugging unused manifold ports as part #05-05 for a set of 

four.

13

12

Содержание ValveLink 8.2

Страница 1: ...ValveLink8 2 User s Manual AutoMate Scientific R E A D Y F O R R E S E A R C H TM...

Страница 2: ...ve Tubing Valve Care Perfusion Pencil Flow Rates Perfusion Pencil Multi barrel Manifold Perfusion Pressure Kit Josef Kewekordes II Heath Lukatch Dave Blau ValveLink8 2 Part 01 18 For indoor use only f...

Страница 3: ...RCA jacks on the back of the box Why not connect valves directly to your computer Several reasons 1 Most computer interfaces provide 5V signals not strong enough to drive most valves 6 to 12V DC and h...

Страница 4: ...wheels and pipette pullers ValveLink8 2 Laboratory Applications Perfusion A ValveLink8 2 can be programmed to precisely switch solutions perfusing over biological samples for physiological research T...

Страница 5: ...on inline heater plus multi channel focal drug delivery maintain bath temperature and rapid drug wash out with a high flow bath line while quickly switching 4 8 preheated solutions through the Perfusi...

Страница 6: ...ld also see the photo on page 5 Find additional photos and diagrams of the pressure system on page 15 2 3 4 1 Assemble ringstand base and rod 2 Attach valve unit 3 Insert syringes with stopcocks Pinch...

Страница 7: ...ine directly to the outflow of the manifold for running cleaning solutions and drying your lines quickly Determine dead volume from the manifold to prep by measuring time and volume required to flush...

Страница 8: ...e Tips reservoir height 40cm Gravity 0 2psi air pressure 0 07ml min 4 6ml hour 8psi air pressure 0 25ml min 15ml hour PTFE Micro manifold Use To use the micro manifold you either need pinch valves or...

Страница 9: ...and removable tips are shipped non sterile They can be chemically sterilized or autoclaved However repeated autoclaving may weaken the adhesive bond between the luer lock fitting and the needle Maint...

Страница 10: ...es mirroring the computer signals low off high on with one millisecond accuracy Usually data acquisition software like pClamp Pulse LabView etc or custom user software is programmed to send digital si...

Страница 11: ...nt panel analog output BNC as an event marker every time a valve opens or closes The marker voltage occurs regardless of how the valve was opened or closed i e manual pushbutton or computer digital in...

Страница 12: ...an electrical path for the spill sensor circuitry In this way a drop of liquid passing the sensor strip appears as a sharp drop in electrical resistance between the two otherwise disconnected copper c...

Страница 13: ...traces A similar conductive epoxy is also available It makes the connection of two wires to your silver traces even easier Simply draw two parallel silver lines approximately 1 16 1 mm apart on any s...

Страница 14: ...between two ValveLink8 2s and control up to 16 valves with a single 4 bit TTL source Set the first ValveLink8 2 to Mode 2 set the second to Mode 3 and connect TTL input bits 1 4 from both ValveLink8 2...

Страница 15: ...ur version of Windows the first time a ValveLink8 2 is installed by USB cable Windows will find new hardware A dialog box will appear asking if you want Windows to connect to Windows Update to search...

Страница 16: ...on rotary switch on the back of each ValveLink8 2 next to its USB port If you connect multiple ValveLink8 2s to the computer be sure to set each one to a different address If you connect a ValveLink8...

Страница 17: ...one valve will be open at a time on all of the networked controllers When the System Status switch is set to Local Control you can click the Mode button on any ValveLink8 2 sub window to access a dia...

Страница 18: ...ar as BNC connectors on the front panel Digital outputs 4 7 are located on the back panel of newer DigiData interfaces Older DigiData 1200 s only have four outputs 2 The Modes chapter explains the two...

Страница 19: ...tubing at an angle and wetting it makes it easier to insert Backflow When a manifold is used in a plumbing arrangement none of its ports can be left unconnected or liquid will simply flow backwards o...

Страница 20: ...g see 1 above and the section below If it qualifies for all three of the above AutoMate Scientific will accept the valve for subsequent return to the factory Repair or replacement is at their discreti...

Страница 21: ...e Safety and Handling data sheets for the above described chemicals are enclosed c We accept full responsibility for any injury to AutoMate Scientific employees or employees of the valves original man...

Страница 22: ...e power requirements of this product 6 Do not allow anything to rest on the power cord Do not locate this product where people will walk on the cord 7 If you have to use an extension cord with this pr...

Страница 23: ...nsor licensees be liable for contingent special direct indirect or consequential damages for the breach of any express or implied warranty or resulting from the use failure or malfunction of any produ...

Страница 24: ...entific Inc 812 Page St Berkeley CA 94710 U S A 800 998 MATE 6283 international 510 845 6283 fax 510 280 3795 e mail info autom8 com www http www autom8 com AutoMate Scientific R E A D Y F O R R E S E...

Отзывы: