Section 8: Maintenance
8.1 Preventive Maintenance
1. Keep the pressure vessel lid bolts lubricated using an anti-seize
agent. Repeated insertion and extraction of stainless steel bolts
in a stainless steel vessel without using anti-seize paste can
cause galling. Lower speed insertion and extraction of bolts can
also help to prevent galling. Use ¼ to ½ the speed for
steel in steel.
2. Periodically lubricate the O-ring with silicone lubricant. The
O-ring is made of synthetic rubber and is located in the pressure
vessel flange. Damaged or worn O-rings should be replaced with
the same material.
8.2 Changing Fuses
1. Control Circuit Fuse: The control circuit fuse is a 2 amp, 250 volt
5 mm x 20 mm BUssmann GMA (or equivalent) fuse. It is located
at the left rear of the system.
2. Heater Circuit Fuse: The heater circuit fuse is a 10 amp, 250
volt semiconductor-type BUssman GMA 2A (or equivalent) fuse.
It is located inside the left control cabinet. If the heater circuit
fuse should malfunction, contact the ATS Service Department at
+1-724-283-1212 for instructions on how to service this fuse.