29
Verify “Age” was
c.
selected from the touch
screen immediately
after loading the
samples, securing the
vessel, and latching the
lid as described in the
previous steps.
If the PAV returns to the setpoint but
ii.
does not pressurize completely, check
that the air line is connected and open
to allow 340-350 psi of airflow.