15
4.4.1 Regulator Connections
1. Secure the air connection to the air tank using the supplied male
and female connections.
2. Connect the hose to the regulator.
NOTE: Do NOT connect the hose to the part marked FLOW. This is
the pressure relief valve, and nothing should be connected to it.
NOTE: Outside of the United States, a different adapter may be
required. A CGA adapter is included.
3. Attach the hose to the PAV using the quick connections. Connect
the male end of the hose to the female connection on the PAV by
sliding the female connection towards the PAV, connecting the
hose, and releasing the slide.
4. Verify that all connections are tightly sealed by applying soapy
water on the connections and turning on the air. If suds appear,
apply more compound or tape to seal the leak.