25
Once the PAV has reached the setpoint, the display
i.
will prompt the operator to insert the specimens,
secure the lid, and select “Age.”
Carefully remove the specimen holder
i.
from the PAV as it will be very hot.
NOTE: The PAV needs to experience heat
loss so it knows what to do after “Age” is
selected. If it does not experience heat
loss, the PAV gets confused and does
not know what to do next. This heat loss
should happen when the specimens are
loaded into the PAV.
Leave the vessel lid open. The PAV is
ii.
designed to recover from the heat loss
during this stage. If the vessel lid is closed
and the temperature doesn’t drop enough,
the PAV may give erratic results and have a
difficult time recovering to the setpoint.
Load specimens into rack.
iii.