13
3. After leveling the ring, hand rotate the horizontal-set screws in
order to secure the ring in the bottom of the vessel.
4.4 Air and Power Connections
1. Install the air pressure line to the 1/4” male pipe-threaded fitting
provided on the back of the control cabinet.
CAUTION: Use a regulated air supply of 340 to 350 psi
(2.34 to 2.41 MPa). Do not exceed 350 psi (2.41 MPa) or
the PAV’s performance may be impaired by excessive
pressure.
2. A connector is provided with the unit. This connector may
be changed if it does not match the available receptacle. Power
requirements are: 230-240 VAC, ±5 V, 10 amp, single-phase, 50/60
Hz.
CAUTION: The PAV requires a minimum of 230 VAC to
operate correctly. Check the supply voltage to ensure it is
230-240 VAC. If the supply voltage is less than
230 VAC, a Step-Up or Buck-Boost transformer can be
supplied to provide the proper operating voltages.
Contact ATS. Be prepared to provide the ATS Service
representative with information about the available AC
voltage so the correct transformer wiring configuration
can be provided.
CAUTION: The PAV and control system must be grounded
according to national and local code requirements.