pH / ORP / ISE / DO / Conductivity Measurement Products Lines
“Sensor Diagnostics” Menu
Sensor diagnostics can be obtained for all connected
sensors that have been properly configured for the
node and sensor type. Simply select the working
channel for the sensor for which you wish to get
diagnostic information and click on the “
Get Sensor
Info
” button. The various fieldx as appropriate will
be populated from the most current data obtained
from the connected sensor. The most current
software revision for each sensor type is shown in
these example sensor diagnostic screenshots.
Date Stamps:
•
Manufacture Date:
When smart digital HiQDT modbus sensor was fabricated at factory
o
Year Manufacture
Range fom 18 for 2018 up to 99 for 2099
o
Month Manufactured Range from 1 for January to 12 for December
o
Date Manufactured
Release date for sensor in year and month of manufacture
§
1-year shelf-life warranty begins from this date
Sensor Statistics:
•
Sensor Serial Number:
Complete traceability of given sensor – Broken up into three separate fields:
o
Serial Number Month Range from 1 for January to 12 for December
o
Serial Number Letter See * note below to convert number shown into alpha character block for sensor
o
Serial Number
Range from 00 to 99 for given alpha character block
•
Type:
pH, standard / wide range ORP or D.O. depending upon connected sensor
•
Software Revision:
Firmware on sensor board (contact factory to ensure most current version)
•
Item Number:
Completely defines all features and capabilities of given sensor
NOTE: Based upon screenshot shown serial number is 18.09-D-27 based upon assemblying the four fields that comprise the complete serial number
Temperature Peak Values:
•
Min Temperature:
Lowest temp (
°
C) experienced by sensor after manufacture date when energized
•
Max Temperature:
Highest temp (
°
C) experienced by sensor after manufacture date when energized
Field Use Time & Dampener Setting:
•
Integral Time Tracking:
The total days the HiQDT sensor has been energized after manufacture date
•
Dampener Delay:
See ** note below to convert number shown into dampener settings in seconds
* 0=A, 1=b, 2=C, 3=d, 4=E, 5=F, 6=g, 7=H, 8=i, 9=J, 10=L, 11=n, 12=o, 13=P, 14=r, 15=S, 16=t, 17=U, 18=Y
** 0=1, 1=2, 2=3, 3=4, 4=5, 5=8, 6=10, 7=15, 8=20, 9=30
Units are Seconds