“Adjust Sensor Dampener” Menu
The available choices for sensor dampener of of the
sensor are shown to the right.
It can be desirable to reduce the dampening time before
performing calibration to expedite the calibration
process.
BE SURE TO RETURN THE DAMPENER BACK TO
THE APPROPRIATE HIGHER VALUES FOR USE IN
CONTINUOUS FIELD USE IF YOU RETURN THE
DAMPENER FOR FASTER CALIBRATIONS.
“Reset All Calibrations” Menu
This will reset ALL available calibrations back to the
values at time of dispatch from the factory.
Once the reset all calibrations call has been performed it
cannot be undone. With this in mind please be sure that
you wish to reset the calibrations before proceeding.
You may be asked to reset the calibrations on your
sensor as part of a troubleshooting in case you have
unusual calibration results.