INSTALLATION
3-1
UNPACKING
Inspect the chamber in its crate for obvious damage that may have occurred during shipment. If
any such damage exists, file a claim with the carrier within 7 days.
If crated, the wooden crate can be dismantled with a hammer and the steel compression straps
can be cut with tin snips. Remove clear plastic wrap from the outside of the chamber. Be
careful not to pull excessively on any protruding switches, pipes etc.
PLACEMENT
Place the chamber on a level, hard surface.
Allow a minimum of 18" clearance in front and rear of chamber for ventilation across air cooled
condenser. The internal components and general performance of the chamber will be impaired if
it is placed in a corner or boxed in by other equipment. Do not block the air vent on the top or
the sides of the chamber because overheating could occur.
POWER REQUIREMENTS
The specific power requirements for the individual chamber are stamped on the model nameplate
affixed to the rear of the chamber. This is the correct requirement for the chamber and should be
followed precisely in order for maximum efficiency and performance standards to be met. The
power must also include a positive earth ground to avoid any shock hazards.
ATTENTION: Not all products from Associated Environmental Systems ship with a means of
disconnect from the main voltage feed. If equipped with a wall plug, this is often sufficient to
satisfy the disconnect requirement of the local electric code. However in instances where the
chamber is hard wired into an electrical service panel, code often requires the installation of a
wall mounted disconnect switch. It is the responsibility of the customer to determine the local
requirements for their chamber. A chamber mounted disconnect switch is available as an option
from Associated Environmental Systems.
INSTALLATION
Observe and follow the instructions given on any tags affixed to the chamber itself. The tags
have specific instructions relating to that individual chamber. The information on the tags may
not be found in this manual.