SECTION VIII A-2000™ SPECIFICATIONS
- 8-3 -
8.1.1
Type of Protection against Electric Shock of the System:
Class 1: Equipment in which protection against electric shock does not rely on basic
insulation only, but which includes an additional safety precaution. Means are provided
for the connection of the equipment to the protective earth conductor in the fixed wiring of
the installation in such a way that accessible metal parts cannot become live in the event
of a failure of the basic insulation.
8.1.2
Degree of Protection against Electric Shock of the System:
Type BF: Equipment providing a degree of protection against electric shock regarding
allowable leakage currents and reliability of the protective earth ground connection with an
F-type applied part. An F-type applied part is isolated from all other parts of the
equipment to such a degree that the patient leakage current allowable in single fault
condition is not exceeded when a voltage equal to 1.1 times the highest rated AC supply
voltage is applied between the applied part and earth.
8.1.3
Degree of Protection against the Ingress of Water:
Monitor degree of protection rating: IPX2 (ingress of water vertically dripping) at up to 15
degree angle.
DSC degree of protection rating: IPX4 (splash proof)
8.1.4
Mode of Operation of the System:
Continuous: Operation under normal load for a normal period, without exceeding the
specified limits of temperature.
8.1.5
Classification:
MEDICAL ELECTRONIC EQUIPMENT
CLASSIFIED BY UNDERWRITERS LABORATORIES INC.®
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH
UL 2601-1, CAN/CSA C22.2 NO. 601.1
8.2
OPTIONS AND ACCESSORIES:
•
BIS Sensor
•
BIS Sensor Patient Interface Cable
•
Sensor Simulator
•
Pole Clamp for Monitor
•
Monitor Stand
•
Printer
Содержание BISPECTRAL INDEX A-2000
Страница 9: ...SECTION I INTRODUCTION 1 4 Figure 1 1 The A 2000 BIS Monitoring System...
Страница 11: ...SECTION I INTRODUCTION 1 6...
Страница 19: ...SECTION III PRINCIPLES OF OPERATION 3 2 Figure 3 1 The A 2000 System Block Diagram...
Страница 20: ...SECTION III PRINCIPLES OF OPERATION 3 3 Figure 3 2 The A 2000 Signal Flow Diagram...
Страница 35: ...SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 6...
Страница 53: ...SECTION VI DIAGNOSTICS AND TROUBLESHOOTING 6 18...
Страница 59: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 6 Flange Figure 7 3 The Keypad Cable Tension Flange...
Страница 61: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 8 Figure 7 4 The Four Mounting Screws...
Страница 71: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 18 L bracket Figure 7 15 The L Bracket Installed...
Страница 76: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 23 Tie Wraps J 300 Figure 7 19 Battery and cable...
Страница 85: ...SECTION VIII A 2000 SPECIFICATIONS 8 4 8 3 Digital Signal Converter DSC Connector LEMO 5 female...
Страница 91: ......