SECTION VI DIAGNOSTICS AND TROUBLESHOOTING
- 6-13 -
6.2.3
Manual Diagnostics
Diagnostic tests may be initiated by selecting them from the Diagnostic Menu.
1. Press the [MENU/EXIT] key from the Main Screen to access the Setup menu.
2. Highlight
Advanced Setup
using the [
▲
] or [
▼
] key, then press [SELECT].
3. Highlight the
Diagnostic Menu
, then press [SELECT]. The Diagnostic Menu
appears.
4. Select the desired option from this menu:
Select:
To:
DSC Self Test
Run the DSC self test diagnostics.
Display Self Test
Run the EL Display self test diagnostic.
Diagnostics Codes ON
Display diagnostics codes in the Message Region of the
screen.
6.2.3.1 DSC Self Test
The DSC self test is a thorough test of the entire signal processing chain including the
signal processing computers. Selecting this item will initiate an extensive DSC test and a
printer test (if one is connected). Do not run this test while electro-cautery equipment is in
use. To begin the test:
1. Highlight
DSC Self Test
, then press [SELECT]. The DSC testing screen will display.
2. The following information appears:
DSC Test Results:
In Progress
Ch1
Ch2
Noise
(µV)
* * *
* * *
Hi-Pass Blocked
(µV)
* * *
* * *
Hi-Pass Normal
(µV)
* * *
* * *
Gain
(µV)
* * *
* * *
The DSC test consists of four sub-tests. The first checks the noise, the second checks
the high-pass filter performance with the amplifiers blocked (fast response), the third
checks the high-pass filter performance with the amplifiers unblocked (high bandwidth),
and the fourth checks the amplifier gains.
The results of each of these sub-tests are displayed for each of the two channels. If any
result is outside the allowed range, the number has an asterisk appended to it. The
allowed range for each of the sub-tests are:
Noise:
0
-
0.24
(µV)
Hi-Pass Blocked:
47.1
-
63.9
(µV)
Hi-Pass Normal:
14.7
-
18.5
(µV)
Gain:
49.2
-
60.2
(µV)
If the Digital Signal Converter is properly connected, the ‘***’ symbols displayed for each
test parameter will change to numeric values for the corresponding channel.
“DSC Test
Results: PASS (or FAIL*)”
message will appear at the end of test. If test fails, check
the connection between the DSC and the monitor, make certain that electro-cautery
equipment is not in use, and run DSC self test again. If the DSC test fails again, the DSC
requires service.
Содержание BISPECTRAL INDEX A-2000
Страница 9: ...SECTION I INTRODUCTION 1 4 Figure 1 1 The A 2000 BIS Monitoring System...
Страница 11: ...SECTION I INTRODUCTION 1 6...
Страница 19: ...SECTION III PRINCIPLES OF OPERATION 3 2 Figure 3 1 The A 2000 System Block Diagram...
Страница 20: ...SECTION III PRINCIPLES OF OPERATION 3 3 Figure 3 2 The A 2000 Signal Flow Diagram...
Страница 35: ...SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 6...
Страница 53: ...SECTION VI DIAGNOSTICS AND TROUBLESHOOTING 6 18...
Страница 59: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 6 Flange Figure 7 3 The Keypad Cable Tension Flange...
Страница 61: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 8 Figure 7 4 The Four Mounting Screws...
Страница 71: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 18 L bracket Figure 7 15 The L Bracket Installed...
Страница 76: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 23 Tie Wraps J 300 Figure 7 19 Battery and cable...
Страница 85: ...SECTION VIII A 2000 SPECIFICATIONS 8 4 8 3 Digital Signal Converter DSC Connector LEMO 5 female...
Страница 91: ......