Ascon Tecnologic - KM2 Line - ENGINEERING MANUAL - PAG. 19
[108] Pr.b3 - Wait band of the third soak
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
.
Range:
OFF... 9999 engineering units.
Note:
For more details see [98] Pr.b1 parameter.
[109] Pr.E3 - Events of the third group
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
.
Range:
00.00... 11.11 where:
0
Event OFF;
1
Event ON.
Note:
For more details see [109] Pr.E1 parameter.
[110] Pr.S4 - Set Point of the fourth soak
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
.
Range:
From
[75] SPLL
to
[76] SPHL
;
oFF
Program end.
Note:
For more details see [100]Pr.S2 parameter.
[111] Pr.G4 - Gradient of the fourth ramp
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
and
[110] Pr.S4 is different from
oFF
.
Range: 0.1... 999.9
engineering units per minute;
inF
Step transfer.
[112] Pr.t4 - Time of the fourth soak
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
and
[110] Pr.S4 is different from
oFF
.
Range: 0.00... 99.59
time units.
[113] Pr.b4 - Wait band of the fourth soak
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
and
[110] Pr.S4 is different from
oFF
.
Range:
From OFF to 9999 engineering units.
Note:
For more details see [98] Pr.b1 parameter.
[114] Pr.E4 - Event of the fourth segment
Available:
When [91] Pr.F is different from
nonE
,
[100] Pr.S2 is different from
oFF
and
[105] Pr.S3 is different from
oFF
and
[110] Pr.S4 is different from
oFF
.
Range:
00.00... 11.11 where:
0
Event OFF;
1
Event ON.
Note:
For more details see [99]Pr.E1 parameter.
[115] Pr.St - Program status
Available:
When [91] Pr.F is different from
nonE
.
Range: run
Program Run;
HoLd
Program Hold;
rES
Program reset.
Note:
This parameter allows to manage program execution
by a parameter.
]
PAn group - Operator HMI
[116] PAS2 - Level 2 password: Limited access level
Available:
Always.
Range: oFF
Level 2 not protected by password
(as level 1 = Operator level);
1... 200
.
[117] PAS3 - Level 3 password:
Complete configuration level
Available:
Always.
Range: 3... 200
.
Note:
Setting [118] PAS2 equal to [119] PAS3, the level 2
will be masked.
[118] uSrb -
button function during RUN TIME
Available:
Always.
Range: nonE
No function;
tunE
Auto-tune enabling. A single press (longer
than 1 s) starts the auto-tune;
oPLo
Manual mode. The first pressure puts the instru-
ment in manual mode (oPLo) while the second
one puts the instrument in Auto mode;
AAc
Alarm acknowledge;
ASi
Alarm reset;
chSP
Sequential Set Point selection (note 1);
St.by
Stand by mode.
The first press puts the instrument in stand
by mode while a second one puts the instru-
ment in Auto mode;
Str.t
Timer run/hold/reset (note 2);
P.run
Program run (note 3);
P.rES
Program reset (note4 );
P.r.H.r
Program run/hold/reset (note 5);
HE.co
Heat using SP1/Cool using SP2 (note 6).
Notes: 1.
When “Sequential Set Point selection” is used,
every press of the
button (longer than 1
second) increase the value of A.SP (active Set
Point) of one step. The selection is cyclic:
SP -> SP2 -> SP3 -> SP4.
When a new Set Point is selected using the
key, the display will show for 2 seconds the
acronym of the new Set Point (e.g. SP2).
When “Sequential Set Point selection” is used, the
number of selectable Set Points is limited by [74] nSP.
2.
When “Timer run/hold/reset” is selected, a short
press starts/stops (hold) timer count while a long
press (longer than 10 second) resets the timer.
3.
When “Program run” is selected, the first press
starts the program execution but a second press
restarts the program execution from the beginning.
4.
When “Program reset” is selected, a short press
allows it to reset the program execution.
5.
When “Program run/hold/reset” is selected, a short
press starts/stop (Hold) program execution while a
long press (longer than 10 s) resets the program.
6.
Setting [118] = HE.co the instrument enables the
parameters [66] tcH, [67] rcG and [68] tcc.