20
Installation
Assembling the XY Autosampler
that the arm locking screw and safety shield have been removed and the
horizontal arm can move.
To install the vertical arm:
1
Remove the cover plate from the front of the horizontal arm by removing its
three screws.
2
Locate the hexagonal-shaped control rod and horizontal slider by looking
down into the horizontal arm. See Figure 5.
3
Using your finger, press on the control rod where it passes through the
horizontal slider. At the same time, pull the needle foot towards the front of
the horizontal arm. This causes the white plastic plug and control rod to move
forward slightly. When the white plastic plug is no longer flush with the front
of the horizontal arm, remove it and the control rod from the horizontal arm.
4
Pull the needle foot, towards the front of the horizontal arm, as far as it will go.
Figure 5
Installing the Vertical Arm—Part 1
5
Position the vertical arm onto the horizontal slider. When viewed from the
front of the XY autosampler, the vertical arm’s needle holder bracket is on the
right.
cover plate
plastic plug
control rod
horizontal slider
needle foot