Page 8 ·
TECPak
User’s Manual
Ventilation
The
TECPak
has vent holes on the front and rear of the unit. You must not block
these vent holes, or overheating may occur, causing damage to the unit.
Rack Mounting
A rack mount kit (
p/n 1402-RM
) for standard 19” racks is available for the
TECPak
, and supports the rack mount of up to five units in either a 2U or 3U
high opening, depending on the orientation of the
TECPak
. Because the units
only ventilate from the front and rear, no air gapping is required above or below
the units, and they can be rack mounted immediately above other equipment
(no rack spacers required).
Warm-up and Environmental Considerations
In order to achieve the highest level of accuracy, the
TECPak
should be
powered on for at least one hour prior to taking measurements. In addition,
ensure that the unit is not operating outside the ambient temperature range or
humidity conditions.
CAUTION
Do not operate the unit above +40°C ambient, and
ensure the instrument is properly ventilated, or the unit
may overheat and possible damage to the instrument
may occur.