Name
Description
Factory
Default
Computer
Command
R-Low Lim
R-Low Lim
is the sensor low
limit. If operating in R Mode and
the actual sensor measurement
drops below this value, the
output will be turned off. R
limits are not monitored in Ite
Mode. Limit monitoring can
also be disabled in software
(see TEC:ENAB:OUTOFF
command).
0.01k
Ω
TEC:LIM:RLO
R-High Lim
R-High Lim
is the sensor high
limit. If operating in R Mode and
the actual sensor measurement
exceeds this value, the output
will be turned off. R limits are
not monitored in Ite Mode. Limit
monitoring can also be
disabled in software (see
TEC:ENAB:OUTOFF
command).
45k
Ω
TEC:LIM:RHI
Tol Time
Tolerance time is the amount of
time, in seconds, that the actual
temperature must be within the
set point tempe/- the
Tol Temp value for the unit to
be considered in tolerance.
5 seconds
TEC:TOL
Tol Temp
Tolerance temperature is a
temperature band (in °C)
around the set point
temperature. When the actual
temperature is within this band
for longer than the Tol Time
setting, then the unit is
considered to be in tolerance.
0.1°C
TEC:TOL
H/C Mode
This selects the heating and/or
cooling mode of the
TECPak
.
See the section below titled
“Resistive Heaters and
Heat/Cool Only Modes” for
more information.
Heat/Cool
TEC:HEATCOOL