background image

4200 Series 

LaserSource

 User’s Manual · Page 15 

 

 

 

Control Modes 

 

 
The 

LaserSource

 offers five laser control modes: 

Io (ACC)

Io HiBW (ACC)

Im 

(AMC)

Po (APC)

, and 

Vf (AVC)

.  

 
Changing the control mode is done through the menu by changing the 

Mode

 

parameter in the menu to one of these values.  
 

Io

 and 

Io HiBW

 modes (referred to collectively as ACC, or automatic current 

control, modes) are used to drive a specific current through the laser diode. 
When in this mode, the set point will be in milliamps, and the 

LaserSource

 will 

drive the desired current through the laser diode as long as the voltage at the 
chosen set point does not exceed the voltage limit. In 

Io

 mode, you will be 

limited to approximately 10Hz modulation. To modulate above that rate, use the 

Io HiBW

, which is a high bandwidth current mode supporting modulation up to 

250kHz (depending on model). 
 

Im

 mode (also referred to as AMC, or automatic monitor photodiode control, 

mode) is used to control the laser diode using the monitor diode feedback. You 
select the target monitor diode current, and the 

LaserSource

 will drive exactly 

enough forward current through the laser diode to generate the selected monitor 
diode current. Only low frequency modulation (10Hz or less) is possible in 

Im

 

mode due to the feedback latencies of the photodiode itself. 
 

Po

 mode (also referred to as APC, or automatic power control, mode) is simply 

Im

 mode with a mathematical constant applied to the set point, providing a 

convenient way of operating in milliwatts. Using the 

PD Response

 factor (in 

μ

A/mW), a 

Po

 set point is internally converted to an equivalent 

Im

 set point by 

the driver, which is then used to control the photodiode feedback. For example, 
if the 

PD Response

 factor was 10, then a set point of 1mW would be the same 

as a set point of 10

μ

A.  

 

Vf

 mode (also referred to as AVC, or automatic voltage control, mode) is used to 

control the voltage driven through the device. Unlike ACC mode, AVC mode 
allows the current to drive to whatever level is necessary to achieve the voltage 
set point, so long as it does not exceed the current limit. As with 

Im

 mode, only 

low frequency modulation (10Hz or less) is possible in 

Vf

 mode. 

 

Modulation 

 
The instrument supports external analog modulation using the modulation BNC 

on the back panel of the instrument. Modulation rates vary by model, so see 
your model’s specification for the maximum modulation rates. Only 

Io HiBW

 

mode supports high speed modulation. All other modes of operation have a 
maximum modulation rate of approximately 10Hz.  

Содержание LaserSource 4205

Страница 1: ......

Страница 2: ...nterface to any voltage from 0 to 5V No more tweaking a potentiometer with a screwdriver and a DMM Simply dial in the voltage you need Unlike other laser diode drivers in its class which use inexpensi...

Страница 3: ...de LaserSource Cable 2m p n 1220 This cable has DB 9 male female connectors for interfacing to the LaserMount or other connectorized fixtures LaserSource Cable 2m Pigtailed p n 1221 This cable has a f...

Страница 4: ...ut important information If at any time any of the following conditions exist or are suspected of existing discontinue use of the unit until it can be inspected by qualified service personnel Visible...

Страница 5: ...nstruments or personnel authorized by Arroyo Instruments Modifications done by non authorized personnel will void the warranty Please see the Service section later in this manual for instructions on h...

Страница 6: ...tion serial number and firmware version number Press the Menu button to enter the menu and using the knob turn to the right until the Io Lim setting is displayed Press the knob to edit the setting and...

Страница 7: ...ly damage the unit Changing the voltage selection requires that you remove the selection tumbler from the IPC Input Power Connector IPC Remove the power cord from the unit Using a small flat blade scr...

Страница 8: ...the AC power cord to the unit You must properly ground the unit by plugging the supplied power cord into a three prong grounded outlet or using a three to two prong adapter and connecting the ground...

Страница 9: ...m and therefore inside the rack housing be sure that the internal rack ambient temperature which will typically be several degrees higher than room ambient does not exceed the unit s operating tempera...

Страница 10: ...he LaserSource menu When in remote mode it acts as a Local button returning the instrument to local mode The large adjustment knob located on the right hand side of the unit is used to change the set...

Страница 11: ...t Displayed Measurements Io Current mA PD current or power and voltage Im PD current A Current and voltage Po PD power mW Current and voltage Vf Voltage V Current and PD current or power The instrumen...

Страница 12: ...e made your change press the knob or Menu button to store the value Pressing the Menu button will store and exit the menu while pressing the knob will store the value but leave you in the menu to make...

Страница 13: ...about limits see the section entitled Hardware and Software Limits Maximum PD Resp This factor is used by the unit to convert from monitor photodiode current into optical power The value is in terms o...

Страница 14: ...et this value to Yes Yes Terminal Mode Terminal mode simply echoes any characters received over the serial or USB interfaces No Msg Term This controls the output message termination and can be set to...

Страница 15: ...or diode current Only low frequency modulation 10Hz or less is possible in Im mode due to the feedback latencies of the photodiode itself Po mode also referred to as APC or automatic power control mod...

Страница 16: ...e via USB is just as simple as using the serial port In fact once you have installed the USB drivers the instrument will appear as a virtual serial port that you can use just like a normal serial port...

Страница 17: ...8 9 Laser Anode Shell Earth Ground Output Connector DB 9 Female Arroyo Instruments has followed industry conventions for laser DB 9 connections and is likely compatible with pin outs from other vendor...

Страница 18: ...on using the USB interface see the Computer Interfacing Manual which is included on the CD that accompanied this product RS232 Connector The RS232 connection is male DB 9 connector wired in a NULL mod...

Страница 19: ...ation of the drive circuit and photodiode measurement circuit Arroyo Instruments carries two cables specifically designed for this application Part number 1220 is a two meter cable designed for use wi...

Страница 20: ...Using Limits The LaserSource provides several limit features for protection of the laser diode These include current voltage photodiode current and photodiode power limits Both the current and voltag...

Страница 21: ...cy reading V 0 05 0 005 Stability ppm time 50 1 hour Temperature Coeff ppm C 100 EXTERNAL MODULATION Input Range 0 10V 10k Modulation Bandwidth 250kHz 250kHz 160kHz MEASUREMENT CHANNELS LASER CURRENT...

Страница 22: ...B Serial only E 501 Interlock shutdown output The interlock input pins 1 and 2 of the output connector were not shorted when the output was on E 504 Laser current limit disabled output The laser outpu...

Страница 23: ...shutdown the output E 516 Incorrect configuration for calibration to start The LaserSource was not configured properly including the mode and output on state to be able to start the desired calibratio...

Страница 24: ...for support Service Service and repair for the LaserSource can be obtained by contacting the distributor from where you purchased the instrument or directly from Arroyo Instruments A complete list of...

Страница 25: ...n which states that the 4200 Series LaserSource Laser Diode Driver is Compliant to the above noted EU Directives and are therefore eligible to bear the CE MARK This Declaration is further validated by...

Страница 26: ...Page 26 4200 Series LaserSource User s Manual Notes...

Страница 27: ...4200 Series LaserSource User s Manual Page 27 Notes...

Страница 28: ...Copyright 2006 Arroyo Instruments All Rights Reserved P N 530 1000 Rev C Copyright 2008 Arroyo Instruments All Rights Reserved P N 530 1000E...

Отзывы: