Page 22 · 4200 Series
LaserSource
User’s Manual
Error Messages
Error
Code
Description
Cause
E-100
General Error
The error code is non-specific, and is
generally used when no other error code is
suitable.
E-102
Message too long
The message is too long to process
(USB/Serial only).
E-123
Path not found
The message used an invalid path
command (USB/Serial only).
E-124
Data mismatch
The message contained data that did not
match the expected format (USB/Serial
only).
E-201
Data out of range
The message attempted to set a value that
was outside the allowable range (USB/Serial
only).
E-202
Invalid data type
When trying to parse the message, the data
was in an invalid format (USB/Serial only).
E-204
Suffix not valid
An invalid number base suffix (radix) was
encountered when parsing a number
(USB/Serial only).
E-501
Interlock shutdown
output
The interlock input (pins 1 and 2 of the
output connector) were not shorted when
the output was on.
E-504
Laser current limit
disabled output.
The laser output was turned off because a
current limit was detected and the
corresponding bit in the OUTOFF register
was set.
E-505
Laser voltage limit
disabled output
The laser voltage exceeded the voltage limit
and the output was turned off.
E-506
Laser photodiode
current limit disabled
output
The laser output was turned off because a
photodiode current limit was detected and
the corresponding bit in the OUTOFF
register was set.
E-507
Laser photodiode
power limit disabled
output
The laser output was turned off because a
photodiode power limit was detected and
the corresponding bit in the OUTOFF
register was set.
E-509
Laser short circuit
disabled output
The laser output was turned off because a
short condition was detected and the
corresponding bit in the OUTOFF register
was set.
Содержание LaserSource 4205
Страница 1: ......
Страница 26: ...Page 26 4200 Series LaserSource User s Manual Notes...
Страница 27: ...4200 Series LaserSource User s Manual Page 27 Notes...