AquaVENT® FD140i, User Manual, English Issue 01
68
Immunity
Immunity test
IEC 60601-1-2
test level
Compliance
level
Guidance to the electromagnec
environment seCng
Electrostac discharge (ESD)
MODE 1
IEC 61000-4-2
± 2 kV, ± 4 kV,
± 6 kV, ±8 kV
contact
± 2 kV, ± 4 kV,
± 6 kV, ±8 kV,
± 15 kV
air
± 4 kV
contact
± 6 kV
air
Floors should be wood, concrete
or ceramic le. If floors are cov-
ered with synthec material, the
relave humidity should be at
least 30%.
At higher levels, during normal
funconality (uninterrupted ther-
apy) it shall be permi@ed a tem-
porary loss of display (blank dis-
play) which may occur due to the
phenomenon of electrostac dis-
charge.
Electrostac discharge (ESD)
MODE 2
IEC 61000-4-2
± 2 kV, ± 4 kV,
± 6 kV, ±8 kV
contact
± 2 kV, ± 4 kV,
± 6 kV, ±8 kV,
± 15 kV
air
± 4 kV
contact
± 8 kV
air
Electrical fast transient/burst
MODE 1
IEC 61000-4-4
± 2 kV
100 kHz burst
frequency
0.75 ms
duraon
± 2 kV
100 kHz burst
frequency
0.75 ms
duraon
The mains power quality should
be that of a typical commercial or
hospital environment.
Permi@ed cking sound may
appear along with display altering,
re-selecon of funcons are
available within 0.01 sec, therapy
mode is uninterrupted.
Surge
IEC 61000-4-5
± 1 kV
differenal
mode
± 2 kV common
mode
± 1 kV
differenal
mode
± 2 kV common
mode
The mains power quality should
be that of a typical commercial or
hospital environment.
Voltage dips, short interrupons and
voltage variaons on power supply
input lines
IEC 61000-4-11
< 5% U
T
(>
95% dip in U
T
)
for ½ cycle
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
< 5% U
T
(> 95%
dip in U
T
) for 5
seconds
< 5% U
T
(>
95% dip in der
U
T
) for ½ cycle
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
< 5% U
T
(> 95%
dip in U
T
) for 5
seconds
The mains power quality should
be that of a typical commercial or
hospital environment.
If the user and/or operator of the
AquaVENT® FD140i requires
connued operaon during power
mains interrupons, it is
recommended that the
AquaVENT® FD140i be powered
from an uninterrupble power
supply (UPS) or a ba@ery.
Magnec field at the power
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnec fields
should be at levels characterisc
of a typical locaon in a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to applicaon of the test level.
Содержание 10190001
Страница 1: ...User Manual English...
Страница 3: ...AquaVENT FD140i User Manual English Issue 01 3 Contents...
Страница 6: ...AquaVENT FD140i User Manual English Issue 01 6 1 Introduc on...
Страница 13: ...AquaVENT FD140i User Manual English Issue 01 13 2 AquaVENT FD140i Overview...
Страница 15: ...AquaVENT FD140i User Manual English Issue 01 15 AquaVENT FD140i Overview 10 11 9 12 14 17 15 13 16...
Страница 20: ...AquaVENT FD140i User Manual English Issue 01 20 Device setup 3 Device Setup...
Страница 33: ...AquaVENT FD140i User Manual English Issue 01 33 4 Using the AquaVENT FD140i...
Страница 49: ...AquaVENT FD140i User Manual English Issue 01 49 5 Alarms and No ca ons...
Страница 58: ...AquaVENT FD140i User Manual English Issue 01 58 6 Maintenance and Repair...
Страница 60: ...AquaVENT FD140i User Manual English Issue 01 60 7 Cleaning and Decontamina on...
Страница 62: ...AquaVENT FD140i User Manual English Issue 01 62 8 Technical Speci ca ons...
Страница 71: ...AquaVENT FD140i User Manual English Issue 01 71 9 European Community Declara on of Conformity...
Страница 73: ...AquaVENT FD140i User Manual English Issue 01 73 10 Disposal...
Страница 75: ...AquaVENT FD140i User Manual English Issue 01 75 11 Appendices...