Alarms and Noficaons
Alarm
Message
Alarm
ID No.
Alarm
Priority
Alarm Condi'on
Correc've Ac'on
Notes
Calibraon
PP Sensor
11
Medium
Paent Pressure sensor
calibraon cannot be
retrieved
Recalibrate device
If recalibraon does not clear the
alarm, return the device to an
approved service centre
Ba@ery
Fault
13
Medium
Ba@ery not fi@ed or
faulty ba@ery installed
Have device serviced by
authorised technician
-
5V
14
Medium
5V supply to device is
greater than +/- 20%
Restart system. If fault
connues, return the
device to an approved
service centre
-
Supply AIR
16
Medium
No AIR supply
connected
Connect AIR supply
When air supply is disconnected,
FiO
2
is automacally set to 100%
Supply O
2
17
Medium
No O
2
supply connected
Connect O
2
supply
When oxygen supply is
disconnected, FiO
2
is
automacally set to 21%
AC Supply
18
Low
AC Supply disconnected
Connect device to AC
supply
Single Alert Alarm
O
2
Sensor
19
Medium
O
2
sensor defect
Have device serviced by
authorised technician
-
Paent
Pressure
Sensor
defect
20
Medium
Device fails to sense AIR
or O
2
Flow and/ or
paent pressure
Have device serviced by
authorised technician
-
Touch Held
21
Low
Touch screen held
for more than 20
seconds
Release touch screen
-
Bu@on Held
22
Low
Front panel bu@on
held for more than 5.5
seconds
Release bu@on
-
Ba@ery
Charge
23
Medium
Ba@ery charge level at
20% or below
Connect device to AC
supply
Single Alert Alarm
Содержание 10190001
Страница 1: ...User Manual English...
Страница 3: ...AquaVENT FD140i User Manual English Issue 01 3 Contents...
Страница 6: ...AquaVENT FD140i User Manual English Issue 01 6 1 Introduc on...
Страница 13: ...AquaVENT FD140i User Manual English Issue 01 13 2 AquaVENT FD140i Overview...
Страница 15: ...AquaVENT FD140i User Manual English Issue 01 15 AquaVENT FD140i Overview 10 11 9 12 14 17 15 13 16...
Страница 20: ...AquaVENT FD140i User Manual English Issue 01 20 Device setup 3 Device Setup...
Страница 33: ...AquaVENT FD140i User Manual English Issue 01 33 4 Using the AquaVENT FD140i...
Страница 49: ...AquaVENT FD140i User Manual English Issue 01 49 5 Alarms and No ca ons...
Страница 58: ...AquaVENT FD140i User Manual English Issue 01 58 6 Maintenance and Repair...
Страница 60: ...AquaVENT FD140i User Manual English Issue 01 60 7 Cleaning and Decontamina on...
Страница 62: ...AquaVENT FD140i User Manual English Issue 01 62 8 Technical Speci ca ons...
Страница 71: ...AquaVENT FD140i User Manual English Issue 01 71 9 European Community Declara on of Conformity...
Страница 73: ...AquaVENT FD140i User Manual English Issue 01 73 10 Disposal...
Страница 75: ...AquaVENT FD140i User Manual English Issue 01 75 11 Appendices...