AquaVENT® FD140i, User Manual, English Issue 01
59
6.1 Repair
AquaVENT
®
FD140i is intended to provide safe and reliable operaon provided that it is operated and
maintained in accordance with the instrucons provided by the manufacturer. If any irregularity becomes
apparent, you should exercise cauon and have the device inspected by an authorised service centre.
6.2 Servicing Schedule
With regular servicing, the expected service life of an AquaVENT
®
FD140i is 10 years. For more
informaon about maintenance, please refer to AquaVENT® FD140i Technical Manual.
The AquaVENT
®
FD140i should be serviced by an authorised Armstrong Medical service centre according
to the following schedule from the first date of use:
Maintenance and Repair
Recommended
service Interval
Conducted by
Every Six months
Armstrong Medical
Replacement /Service of AEC0355 Oxygen sensor if it is used
Every year
Armstrong Medical
Technical safety inspecon of AquaVENT® FD140i
Every year
Armstrong Medical
Replace Cone Filters in the NIST fiCngs -N2185/06
Every year
Armstrong Medical
Replace the Free Breathing Valve – AEC1222
Every year
Armstrong Medical
Replace the O ring at the 1/8”screw joint of the NIST fiCngs
Every year
Armstrong medical
Replace the O-ring at the Nebuliser gas outlet port
Every year
Armstrong Medical
Replace the O-ring at the Paent pressure gas inlet port
Every year
Armstrong Medical
Replace the O-ring at the 22mm gas outlet port
Every Two years
Armstrong Medical
Replace the Sintered Flow Discs within the manifold -AEC1221
Every Two years
Armstrong Medical
Replace the device ba@ery
Every Six Years
Armstrong Medical
Replace the O
2
pressure regulator
Every Six Years
Armstrong Medical
Replace the Air pressure regulator
Every Six Years
Armstrong Medical
Replace the 2 proporonal (flow sensor) valves
Every Ten Years
Armstrong Medical
It is necessary to replace all acve internal components of AquaVENT® FD140i
for safety reasons during maintenance aGer 10 years.
Instrucons
Every Three years
Armstrong Medical
Replace the Real Time Clock ba@ery
AquaVENT
®
FD140i Servicing schedule from the date of first use.
Содержание 10190001
Страница 1: ...User Manual English...
Страница 3: ...AquaVENT FD140i User Manual English Issue 01 3 Contents...
Страница 6: ...AquaVENT FD140i User Manual English Issue 01 6 1 Introduc on...
Страница 13: ...AquaVENT FD140i User Manual English Issue 01 13 2 AquaVENT FD140i Overview...
Страница 15: ...AquaVENT FD140i User Manual English Issue 01 15 AquaVENT FD140i Overview 10 11 9 12 14 17 15 13 16...
Страница 20: ...AquaVENT FD140i User Manual English Issue 01 20 Device setup 3 Device Setup...
Страница 33: ...AquaVENT FD140i User Manual English Issue 01 33 4 Using the AquaVENT FD140i...
Страница 49: ...AquaVENT FD140i User Manual English Issue 01 49 5 Alarms and No ca ons...
Страница 58: ...AquaVENT FD140i User Manual English Issue 01 58 6 Maintenance and Repair...
Страница 60: ...AquaVENT FD140i User Manual English Issue 01 60 7 Cleaning and Decontamina on...
Страница 62: ...AquaVENT FD140i User Manual English Issue 01 62 8 Technical Speci ca ons...
Страница 71: ...AquaVENT FD140i User Manual English Issue 01 71 9 European Community Declara on of Conformity...
Страница 73: ...AquaVENT FD140i User Manual English Issue 01 73 10 Disposal...
Страница 75: ...AquaVENT FD140i User Manual English Issue 01 75 11 Appendices...