AquaVENT® FD140i, User Manual, English Issue 01
56
Alarm
Message
Alarm
ID No.
Alarm
Priority
Alarm Condi'on
Correc've Ac'on
Notes
O
2
Calibraon
24
Medium
O
2
sensor not calibrated
Calibrate oxygen sensors;
refer to secon 4.5 “Oxygen
sensors calibraon”
Single Alert Alarm
Following calibraon of O
2
sensor, If the O
2
sensor fails at
start up the alarm will be
acvated once
P Min
25
Medium
Paent pressure is less
than applied P Min
alarm limit
Evaluate low pressure
alarm seCngs and increase
pressure seCng if
appropriate
P Min alarm does not acvate
in HFOT or POINT
P Max
26
Medium
Paent pressure is
more than applied
P Max alarm limit
Evaluate high pressure
alarm seCngs and reduce
pressure seCng if
appropriate
P Max alarm does not acvate
in HFOT or POINT
Apnoea
27
Medium
No respiraon detected
for a period greater
than 12 seconds
Check paent and evaluate
apnoea alarm seCngs
-
F Max
28
Low
Respiratory rate is
greater than applied
F Max alarm limit
Evaluate F Max alarm
seCngs and increase F Max
alarm value if appropriate
-
P Limit
29
Medium
Paent pressure
greater than 25cmH
2
O
for CPAP, CPAP Paed
and Helmet
Paent pressure
greater than 15cmH
2
O
for BUBBLE PAP
Evaluate High Pressure
Alarm seCngs and reduce
pressure seCng if
appropriate
P Limit alarm does not acvate
in HFOT or POINT
FiO
2
High
30
Low
Detected FiO
2
level is at
least 5% higher than
the set value
Wait 10 seconds, if alarm
clears proceed with use.
If alarm does not clear
calibrate oxygen sensors;
refer to secon 4.5 “Oxygen
sensors calibraon”. If
calibraon does not clear
the alarm, return the device
to an approved service
centre
Following adjustment of O
2
percentage, FiO
2
high alarm is
deacvated for 30 seconds
FiO
2
Low
31
Low
Detected FiO
2
level is at
least 5% below the set
value or less than 18%
Wait 10 seconds, if alarm
clears proceed with use.
If alarm does not clear
calibrate oxygen sensors;
refer to secon 4.5 “Oxygen
sensors calibraon”. If
calibraon does not clear
the alarm, return the device
to an approved service
centre
Following adjustment of O
2
percentage, FiO
2
low alarm is
deacvated for 30 seconds
Fan Defect
34
Medium
Technical failure
Have device serviced by
authorised technician
-
NOTE: Single Alert Alarms do not repeat audio alarm aGer alarm mute has been acvated.
Alarms and Noficaons
Содержание 10190001
Страница 1: ...User Manual English...
Страница 3: ...AquaVENT FD140i User Manual English Issue 01 3 Contents...
Страница 6: ...AquaVENT FD140i User Manual English Issue 01 6 1 Introduc on...
Страница 13: ...AquaVENT FD140i User Manual English Issue 01 13 2 AquaVENT FD140i Overview...
Страница 15: ...AquaVENT FD140i User Manual English Issue 01 15 AquaVENT FD140i Overview 10 11 9 12 14 17 15 13 16...
Страница 20: ...AquaVENT FD140i User Manual English Issue 01 20 Device setup 3 Device Setup...
Страница 33: ...AquaVENT FD140i User Manual English Issue 01 33 4 Using the AquaVENT FD140i...
Страница 49: ...AquaVENT FD140i User Manual English Issue 01 49 5 Alarms and No ca ons...
Страница 58: ...AquaVENT FD140i User Manual English Issue 01 58 6 Maintenance and Repair...
Страница 60: ...AquaVENT FD140i User Manual English Issue 01 60 7 Cleaning and Decontamina on...
Страница 62: ...AquaVENT FD140i User Manual English Issue 01 62 8 Technical Speci ca ons...
Страница 71: ...AquaVENT FD140i User Manual English Issue 01 71 9 European Community Declara on of Conformity...
Страница 73: ...AquaVENT FD140i User Manual English Issue 01 73 10 Disposal...
Страница 75: ...AquaVENT FD140i User Manual English Issue 01 75 11 Appendices...