12
SECTION
B
Step 4
After the symbol appears on the display, apply the solution to the narrow
edge of the test strip until the meter beeps. Make sure the confirmation
window fills completely.
NOTE
• The meter will switch off if the control solution sample is not applied within 2 minutes of the
symbol appearing on the screen. If the meter turns off, remove the test strip, reinsert, and
start from step 2.
Step 5
A test result will appear after the meter counts down from 5 to 1.
After your control solution result appears on the display, press
for 3 seconds until “check” appears on the display.
When “check” is displayed, the result is stored in the meter’s memory
but it is not included in the averages.
Compare the result displayed on the meter to the range printed
on the test strip bottle or box of individually wrapped test strips.
If the result falls within the specified range, you are ready to use
the meter and test strips to test with blood.
Step 6
Remove the used test strip from the meter’s test strip port. The meter
will turn off automatically. Used test strips should be safely discarded in
disposable containers.
CAUTION
• The ranges printed on the test strip bottle or box of individually wrapped test strips are for the
Assure® Prism Control Solutions only. They do not have any connection to blood glucose level.
NOTE
• Assure Prism Control Solution can be purchased from your local supplier or pharmacy. If they do
not stock it, contact ARKRAY Technical Customer Service at 800.818.8877, option 5, 24 hours a
day, 7 days a week.
Control Solution Range
Control 1:
Control 2: