Installation
The bed should be unoccupied; if this is not possible, take care not to disturb
the patient during installation.
Make sure the mattress platform is flat and level. It is not necessary to remove
the mattress.
Fully undo the fixing clamp screw (C). Position the Egress Assist Rail, in its
lowered position, on the side frame of the bed. Hook the fixing bracket over
the top of the bed frame.
The spacer bar (B) must fit behind the edge of the head end panel (A) as
shown. This will make sure the Egress Assist Rail is correctly positioned on
the bed.
Tighten the fixing clamp screw (C) and make sure the groove in the clamp
block (D) sits against the inner bottom edge of the bed frame.
Repeat this procedure to attach the other Egress Assist Rail on the opposite
side of the bed. Make sure both fixing clamp screws are fully tightened.