27
Caregiver Obligations
Caregiver obligations shall be carried out by personnel with
sufficient
Bolero
knowledge following the instructions in this
IFU
.
Every day
•
Disinfect.
The
Bolero
lift bath trolley has to be disinfected
immediately after every use. ArjoHuntleigh disinfectants
should be used in recommended concentrations as labelled
on the disinfection bottle. See section
Cleaning and
Disinfection Instructions on page 21.
Every week
•
Visually check all exposed parts
, especially where personal
contact is made by either the patient or caregiver. Make sure
no cracks or sharp edges have developed that could cause
the patient or user injury or that has become
unhygienic. Replace damaged parts.
•
Visually check the mechanical attachments
by placing a
foot on the chassis and pull the stretcher upwards with both
hands.No gaps are allowed. Check that all screws are
tightened. Check that the cover over the backrest adjustment
catch is in good condition and properly fastened. Check the
backrest retainer mechanism on both sides.
(See Fig. 1)
•
Check hand control and cable
: Visually check the condition
of the hand control and its cable.
(See Fig. 2)
•
Check that the castors
are properly fixed and are rolling and
swivelling freely. Clean with water (the function can be
affected by soap, hair, dust and chemicals from floor
cleaning).
(See Fig. 3)
•
Check the brakes
on the castors and check the castor brake
lever. Replace damaged parts.
•
Check the mattress/pillow
to make sure there are no cracks
or tears that would allow water to get into the filling. If there
is such damage, replace damaged parts.
(See Fig. 4)
Continue with the steps on the next page.
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Содержание Bolero