13

7.

Decontamination

The following processes are recommended, but should be adapted to comply with 
the local or national guidelines (Decontamination of Medical Devices) which may 
apply within the Healthcare Facility or the country of use. If you are uncertain, you 
should seek advice from your local Infection Control Specialist.

The 

Bari-Breeze

 system should be routinely decontaminated between patients and at 

regular intervals while in use; as is good practice for all reusable medical devices.

During Use

It is recommended that were practical, all parts of the 
system (including mattress, top cover, pump, and 
tubeset) should be cleaned weekly whilst in use.

To clean

Wipe all exposed surfaces with a cloth which has been 
moistened in simple detergent and water. Ensure to 
remove any organic debris that may be adherent to the 
surfaces. Dry.

To disinfect

Wipe all 

cleaned 

surfaces with a solution which 

contains a chlorine-releasing agent diluted up to a 
strength of 10,000 parts per million (depending on local 
policy): an example is 

sodium hypochlorite

. Dry.

Alternatively use a cloth moistened with alcohol (70%) 
and allow to air dry.

Laundering

The mattress cover can be easily unzipped for complete 
removal to allow laundering at 65°C for 10 minutes or 
80°C for 3 minutes to achieve thermal disinfection. This 
complies with HSG (95) 18 Hospital Laundry 
arrangements for used and infected linen (UK) 1995.

DO NOT TUMBLE DRY COVERS ABOVE 30°C.

WARNING

Remove the power cord from the mains supply before cleaning and
decontamination.

Caution

Do not use Phenol-based solutions or abrasive compounds or pads
during the decontamination process as these will damage the
surface coating. Do not boil or autoclave the cover.
Avoid immersing electrical parts in water during the cleaning
process.

Содержание Bari-Breeze

Страница 1: ...Low Air Loss Turn Assist System User Manual Bari Breeze ...

Страница 2: ......

Страница 3: ...llation and Setup 6 Warnings 6 Installation 6 CPR Facility 8 Controls Alarms and Indicators 9 Controls and Indicators 9 Alarms 10 Operation 11 Initial Power Up 11 Comfort Controls 11 Static Mode 11 Turn Options 12 Turn Times 12 To Deflate and Store the Bari Breeze Mattress 12 Decontamination 13 Troubleshooting Guide 14 Spare Parts 16 Warranty and Service 17 Technical Data 18 ...

Страница 4: ... ii ...

Страница 5: ...e mains socket before cleaning and inspection Ensure the system is clean and dry prior to storage Never use sharp objects or electrically heated under blankets on or under the system Store the pump and mattress in the protective bags supplied Only the pump and mattress combination as indicated by ArjoHuntleigh should be used The correct function of the product cannot be guaranteed if the incorrect...

Страница 6: ...lity EMC This product complies with the requirements of applicable EMC Standards The use of accessories not specified by the manufacturer may result in increased emissions by or decreased immunity of the equipment affecting its performance ...

Страница 7: ...ity of the support surface in contact with the patient The Bari Breeze pump is fully programmable and provides the following modes of operation Static Mode Default When Turn therapy settings are set to the system remains flat providing a continuous low pressure surface Turn Rotation Mode Provides an angled support surface to assist with nursing procedures Firm Mode Maxflow Provides a firm support ...

Страница 8: ...2 Bari Breeze Mattress A self contained mattress replacement system with easily detachable components for cleaning Cover Pump Connector Safety Air Pad Plug ...

Страница 9: ... a double zipped cover The top and sides of the cover are constructed from a two way stretch and water resistant material which is vapour permeable for maximum patient comfort The base of the cover is constructed from a tough abrasion resistant material and is fitted with bed attachment straps Safety Air Pad A 50 mm 2 thick sub mattress made of a single air filled cell enclosed in the base to supp...

Страница 10: ... pump from the bed foot rail Keypad The keypad is angled to provide the user with a clear visual display An LED display and individually illuminated control buttons give precise indications of the status of the system at any point in time LED Display Highly visible and bright 7 segment blue LED digital display panel ...

Страница 11: ...d be obtained from the appropriate physician before using the Bari Breeze system A risk assessment tool combined with clinical judgement should be used when determining a patient s level of risk of developing pressure ulcers Patient risk assessment should be an ongoing process as changes in the patient s condition may increase or decrease their risk level The above are guidelines only and should n...

Страница 12: ...ess is placed directly on the bed frame Make sure that the tubeset end of the mattress is towards the foot end of the bed WARNING The Bari Breeze system must be installed on bed frames that are equipped with Safety Sides Please raise safety sides on the bed and lock them in position after the patient is on the mattress NEVER LEAVE A PATIENT UNATTENDED ON THE MATTRESS SYSTEM WITH THE SAFETY SIDES I...

Страница 13: ...or in such a manner that it is a hazard 5 Plug the power cord into the pump and press it in place Uncoil the power cord and plug the cord into a properly grounded AC power source 6 Securely attach the pump connector onto the pump until it clicks and make sure the side bolster and safety air pad plug is secure Care should be taken such that the mattress hose is freely suspended without being pinche...

Страница 14: ...r and simultaneously pull the hose away from the pump To reinflate the mattress 1 Switch the pump into standby mode and reconnect the tubeset and safety air pad plug if removed Switch the pump on and reset the therapy settings IMPORTANT In the event of cardiac arrest depress the two release catches on the tubeset pump connector and at the same time pull the tubeset out and away from the pump ...

Страница 15: ...level will revert to the previous setting and remain in Static mode Turn Options The button selects the turn settings of the mattress Each press of the button selects the next setting For instance to change the mattress from Bi lateral Turn to Static position press the button twice until is selected In Left Turn mode the right air cushions in the mattress will be held at a constant high pressure a...

Страница 16: ...ch can can be locked out to prevent unintentional change of modality cell pressures To activate press the lock button for approximately 2 3 seconds until the blue LED illuminates To deactivate press and hold the lock out button until the LED turns off Alarms Power Fail In the event of a power fail situation the pump will alert the carer by flashing the amber Power Fail LED and sounding the buzzer ...

Страница 17: ...mum of 4 finger width clearance between the bottom of the patient and the safety air pad Recommended Pressure Settings For extra firm support during patient handling or nursing procedures it is recommended to set the mattress pressure to maximum by pressing MAX FLOW When the patient is to be placed in the gatched position it is also recommended to press the button to ensure the patient does not bo...

Страница 18: ...l Turn To enter bi lateral Turn mode press the button until is indicated this is confirmed by the illuminated blue LED Selecting Max Flow will cancel any Turn options set resulting in the system returning to the Static setting Turn Times To set therapy turn times press the button until the desired time is indicated this is confirmed by the illuminated blue LED Turn times are only considered when t...

Страница 19: ...c debris that may be adherent to the surfaces Dry To disinfect Wipe all cleaned surfaces with a solution which contains a chlorine releasing agent diluted up to a strength of 10 000 parts per million depending on local policy an example is sodium hypochlorite Dry Alternatively use a cloth moistened with alcohol 70 and allow to air dry Laundering The mattress cover can be easily unzipped for comple...

Страница 20: ... Hose set connection not correctly fitted Check to ensure the hose connector clicks into place Re connect hose connection to pump ensuring firm fitment Bolster Sub Mattress plug not fitted correctly or failing Check plug is fitted correctly in hose manifold Check integrity of plug Ensure plug is firmly located Replace if neccesary Mattress inflates but doesn t appear to turn No turn angle set Ther...

Страница 21: ...in tubeset connector Reconnect if possible otherwise call Service Engineer Torn or damaged cell Check Call service engineer to replace damaged cell Blower motor inside pump sounds excessively noisy during continuous use The blower inside the pump unit may be approaching the end of its life Check to ensure the mattress is fully inflated Call service engineer to request a replacement pump unit if re...

Страница 22: ... shown below Pump Order Code Description 647301 UK Bari Breeze Pump 647303 USA Bari Breeze Pump 647304 Euro Bari Breeze Pump 647309 ROW Bari Breeze Pump Mattresses Order Code Description 647311 Bari Breeze Mattress 84 x 42 Mattress Covers Order Code Description 647321 Bari Breeze Mattress Cover 84 x 42 ...

Страница 23: ... statutory right of the consumer For service maintenance and any questions regarding this or any other ArjoHuntleigh product please contact For the UK Huntleigh Healthcare Ltd 310 312 Dallow Road Luton Bedfordshire LU1 1TD Tel 44 0 1582 413104 Fax 44 0 1582 459100 For the USA Huntleigh Healthcare L L C 40 Christopher Way Eatontown New Jersey 07724 3327 Tel 800 223 1218 Fax 1 732 578 9889 ...

Страница 24: ...1 Bari Breeze Mattress Size Inflated Width 106cm 42 Height 29cm 11 Length 213cm 84 Weight 9kg 20lb Electrical Safety Standards Complies with EN 60601 1 1990 A13 1996 UL60601 1 1st Edition CAN CSA C22 2 No 601 1 M90 Medical Equipment Degree of protection against electric shock Class I Type BF Degree of protection of ingress of fluids Ordinary IPx0 not protected Mode of operation Continuous ...

Страница 25: ...mental Conditions Operating Temperature range 10 C to 40 C Relative humidity 30 to 75 non condensing Atmospheric pressure 700 hPa to 1060 hPa Storage Temperature range 40 C to 70 C Relative humidity 10 to 100 Atmospheric pressure 500 hPa to 1060 hPa Environmental Protection Please dispose of this pump in accordance with local regulations ...

Страница 26: ...20 ...

Страница 27: ...lerslie Auckland New Zealand T 64 9 525 2488 F 64 9 525 2433 Deutschland HNE Healthcare GmbH Im Hülsenfeld 19 D 40721 Hilden T 49 0 2103 97 11 0 F 49 0 2103 97 11 80 South Africa Huntleigh Africa Pty Ltd Willem Cruywagen Avenue Stand 120 Klerksoord Pretoria South Africa T 27 12 542 4680 F 27 12 542 4982 Danmark Huntleigh Healthcare A S Vassingerødvej 52 DK 3540 Lynge T 45 4 913 8486 F 45 4 913 848...

Страница 28: ... Huntleigh Healthcare Limited 2007 647900EN_01 ...

Отзывы: