11
6.
Operation
Before using the
Bari-Breeze
mattress system, make sure it has been installed
correctly in accordance with “Installation and Setup” on page 6, and ensure that the
side bolster and base plug is secure in the connector at the foot end of the mattress.
Initial Power Up
Press the power/standby button to turn the pump on. An
audible beep will sound and the current comfort level,
turn time, turn mode, and turn angle will be displayed.
These may reflect the therapy settings that were applied last time the system
was in use.
Comfort Controls
Once the mattress is fully inflated and with the patient in
position, adjust the comfort controls to the desired level.
Perform a four finger check as follows:
Place four fingers between the air cushions directly
underneath the sacral region of the patient's body.
There should be a minimum of 4 finger width clearance between the bottom
of the patient and the safety air pad.
Recommended Pressure Settings
For extra firm support during patient handling or nursing
procedures, it is recommended to set the mattress
pressure to maximum by pressing
MAX FLOW (
).
When the patient is to be placed in the gatched position,
it is also recommended to press the
button to ensure
the patient does not bottom out.
In the case of a patient who has an above average weight
to height ratio, it is recommended to
increase
the
pressure/comfort level accordingly.
Static Mode
Max Flow
Pressing the
button sets the blower to maximum
flow, fully inflating the mattress overriding the comfort
control. This feature is only available in static mode. If
pressed during
Turn
mode, it will revert the system to
Static
mode and return the patient to a central position.
Comfort settings however, are retained.
Max Flow
inflates the mattress to maximum pressure for
15 minutes, and this will be indicated by a blue LED and
an ‘
F
’ being shown in the LED display.