10
Turn Times
The time selected for the turn is the amount of time (in
minutes) the mattress will remain in each position
during the selected turn sequence.
Turn Angle
Four settings are available to select the amount of turn
given to the patient. These are selected using the turn
angle button
, where 4/4 is equal to approximately 40
degrees.
Comfort Control Level
These controls adjust the firmness of the mattress. Using
the
button reduces the pressure setting, and the
button increases it.
The current comfort level (1-9) will be displayed in the
LED display.
Lock Out
Pump functions (including power/standby switch) can
can be locked out to prevent unintentional change of
modality / cell pressures. To activate, press the lock
button for approximately 2-3 seconds until the blue LED
illuminates. To deactivate, press and hold the lock out
button until the LED turns off.
Alarms
Power Fail
In the event of a power fail situation, the pump will alert
the carer by flashing the amber
Power Fail
LED and
sounding the buzzer. Once the power is restored to the
pump the alarm will cease and the pump will return to its
previous operating settings.
Low Pressure
In the event of hose disconnection, an ‘
L’
and ‘
P’
will
alternate on the LED display (representing "Low
Pressure") while the buzzer sounds to alert the carer to
the alarm condition. Once the hose is reconnected, the
alarm will cease and the pump resume its previous
therapy settings.