19
1. Stop the therapy and put the pump into
Standby
.
2. Disconnect the tubeset from the pump (refer to
“Disconnecting the Tubeset” on page 10).
3. Activate the CPR control at the head end of the
mattress to deflate it (refer to “To Activate the
CPR Control” on page 20).
To Store the Mattress
Following deflation:
1. Start rolling the mattress from the foot end towards
the head end; stop after
ONE
turn of the mattress.
2. Bring the tubeset connector over the mattress and
secure.
3. Continue to roll the mattress from the foot end
towards the head end.
Make sure the mattress is dry before rolling it up.
Transport Mode
To transport a patient who is lying on the
Alpha
Response
mattress:
1. Stop the therapy and put the pump into
Standby
.
2. Disconnect the tubeset from the pump (refer to
“Disconnecting the Tubeset” on page 10). This will
automatically put the mattress into transport mode.
3. Put the tubeset connector through the transportation
loop to prevent any damage to the tubeset while the
bed is being moved.
The patient will remain supported for up to eight
hours on the mattress.
1
2
3