22
6. Seat Cushion - Pump Operation
General
Refer to the "Initial Inflation" section in
Section 5, Page 16 “Mattress - Pump Operation”.
Seat Cushion
Operation
1. At the start of the inflation sequence, the default
seat-inflation screen is displayed.
• The default patient weight range setting is
Normal
, as indicated by the three
Weight
Range
icons.
• The only cell inflation mode is
Active
(Alternating)
, as indicated by the
Mattress/Seat
Mode
icon.
• For the duration of the inflation sequence, the
patient body is replaced by an “egg timer” icon.
2. When the seat cushion inflation is complete, the
default seat-operating screen is displayed; the “egg
timer” is replaced by the patient body icon.
3. The patient can now sit on the seat cushion.
4. To change the
Weight Range
setting, refer to
Section 4, Page 11 “Controls, Alarms and
Indicators”.
This setting can also be changed while the seat
cushion is being inflated.
To Deflate the Seat
Cushion
To deflate the seat cushion, do the following:
1. Press and hold the
Run/Standby
button for three
seconds to put the pump into
Standby
.
WARNING
Do not place the patient on the seat cushion until it is fully inflated