GENERAL SAFETY

Before you connect the system pump to a mains socket, read carefully all the installation
instructions contained within this manual.
The system has been designed to comply with regulatory safety standards including:
• EN60601-1:1990/A13:1996 and IEC 60601-1:1988/A2:1995
• UL60601-1, UL2601-1 and CAN/CSA C22.2 No. 601.1-M90
• EN60601-1:2006 and IEC 60601-1:2005
• AAMI/ANSI ES60601-1:2006 and CAN/CSA C22.2 No.60601.1(2008)

Safety Warnings

• It is the responsibility of the care giver to ensure that the user can use this product

safely.

• Whilst the patient is unattended, the decision to use safety sides should be used

based on clinical assessment and in line with local policy.

• Alignment of the bed frame, safety sides and the mattress should leave no gap

wide enough to entrap a patient's head or body, or to allow egress to occur in a
hazardous manner where entanglement with the mains power cable and tubeset or
air hoses may result. Care should be exercised to prevent occurrence of gaps by
compression or movement of the mattress. Death or serious injury may occur.

• Make sure that the mains power cable and tubeset or air hoses are positioned to

avoid causing a trip or other hazard, and are clear of moving bed mechanisms or
other possible entrapment areas. Where cable management flaps are provided
along the sides of the mattress, these should be used to cover the mains power
cable.

• Electrical equipment may be hazardous if misused. There are no user-serviceable

parts inside the pump. The pump's case must only be removed by authorised
technical personnel. No modification of this equipment is allowed.

• The mains power socket/plug must be accessible at all times. To disconnect the

pump completely from the electricity supply, remove the plug from the mains
power socket.

• The CPR control and/or the CPR indicator tag must be visible and accessible at all

times.

• Disconnect the pump from the mains power socket before cleaning and inspecting.
• Keep the pump away from sources of liquids and do not immerse in water.
• Do not use the pump in the presence of uncontained flammable liquids or gasses.
• The cover of this product is vapour permeable but not air permeable and may

present a suffocation risk.

• There is no transport mode on the Alpha Response seat cushion.
• Only the pump and mattress or seat combination as indicated by ArjoHuntleigh

should be used. The correct function of the product cannot be guaranteed if
incorrect pump and mattress or seat combinations are used.

Precautions

For your own safety and the safety of the equipment, always take the following
precautions:

• Placing extra layers between the patient and the mattress potentially reduces the benefits 

provided by the mattress and should be avoided or kept to a minimum. As part of sensible 
pressure area care, it is advisable to avoid wearing clothing which may cause areas of 

Содержание Alpha Response

Страница 1: ...Alpha Response with people in mind 464934US_03 01 2016 Instructions for Use...

Страница 2: ...for Use Design Policy and Copyright and are trademarks belonging to the ArjoHuntleigh group of companies ArjoHuntleigh AB 2015 As our policy is one of continuous improvement we reserve the right to mo...

Страница 3: ...the Alpha Response Seat cushion 7 Installing the Pump 8 Controls Alarms and Indicators 11 Control Panel 11 Alarms 15 Mattress Pump Operation 16 Initial Inflation 16 Mattress Operation 17 Raising the...

Страница 4: ...ii Troubleshooting 27 Technical Description 28...

Страница 5: ...des of the mattress these should be used to cover the mains power cable Electrical equipment may be hazardous if misused There are no user serviceable parts inside the pump The pump s case must only b...

Страница 6: ...ecified by the manufacturer may result in increased emissions by or decreased immunity of the equipment affecting its performance Portable and mobile radio frequency RF communications equipment e g mo...

Страница 7: ...on system designed to complement pressure ulcer treatment and prevention protocols The product offers two therapeutic modes Active1 Alternating mode which periodically redistributes pressure away from...

Страница 8: ...mbent position the clinician or patient should select the Backrest Raised mode on the pump which will increase the pressure in the mattress cells to provide optimal pressure redistribution to the pati...

Страница 9: ...3 Static Head Cells CPR 14 Standard Cells Tubeset Control Transportation Loop Cable Management Flap Securing Straps HEAD END FOOT END Detachable Cover Active Alternating Cells Base Cover Fixing Strap...

Страница 10: ...ctures or conditions which may be complicated by a soft or moving surface advice should be sought from an appropriate clinician before use While the Alpha Response systems have been designed to manage...

Страница 11: ...the bed and the CPR at the head end The mattress cover must be uppermost 2 Attach the mattress to the bed frame using the four strap assemblies on the underside of the mattress replacement The strap a...

Страница 12: ...anchor points on the bottom of the mattress replacement This allows for attaching the mattress to different types of bed frame If the bed has divided sections for independent elevation of a patient s...

Страница 13: ...The cells in the seat cushion are in a horizontal position across the chair with the V shape pointing towards the front 3 Secure the seat cushion to the chair by using the fixing straps as shown in th...

Страница 14: ...ty will rely on the anti slip base material of the seat cushion 7 Place the protective cover over the seat cushion and ensure that the logo and the orientation icon printed on the cover are uppermost...

Страница 15: ...ows 1 Locate one of the cable management flaps 2 If necessary open the press studs along the flap 3 Run the mains power cable along the side of the mattress securing the flap round the cable using the...

Страница 16: ...rom the pump This will put the mattress into Transport Mode and will not deflate the mattress To deflate the mattress refer to To Deflate the Mattress on page 18 There is no transport mode on the Alph...

Страница 17: ...to amber After you put the pump in Standby if you press the Run Standby button within approximately 15 seconds the pump goes straight to the Run mode and continues the previous therapy if you wait mor...

Страница 18: ...eight Range or Mode is being changed the Mode Selection icon changes to a tick and flash Pressing the Select button under the tick confirms the new selection If the Select button is not pressed for fi...

Страница 19: ...ines Patient comfort and clinical judgement should be used to select the correct weight range setting To change the Weight Range do the following 1 Press the Weight Mode Select buttons to highlight th...

Страница 20: ...highlight the new Mode icon The new Mode icon is solid with a square border 4 The new Mode icon and square flashes and the Mode Selection icon changes 5 Press the Select button to confirm the new Mode...

Страница 21: ...its target pressure the Activity Indicator stops flashing and starts rotating again Alarms 1 When the pump detects an alarm condition The corresponding visual alarm is displayed on the LCD screen as d...

Страница 22: ...r is illuminated amber to indicate that external power is applied to the pump but the pump is still in Standby The LCD screen is blank 3 Press the Run Standby button to put the pump into the Run mode...

Страница 23: ...on sequence the patient body is replaced by an egg timer icon 3 When the mattress inflation is complete the default mattress operating screen is displayed The egg timer is replaced by the patient body...

Страница 24: ...screen below shows the mattress in Active Alternating mode when the backrest is raised 1 When the bed frame is put back to the backrest horizontal position the clinician or patient should manually sel...

Страница 25: ...over the mattress and secure 3 Continue to roll the mattress from the foot end towards the head end Make sure the mattress is dry before rolling it up Transport Mode To transport a patient who is lyi...

Страница 26: ...rol Use a quick firm pull on the CPR tag to remove the CPR plug The air is rapidly evacuated from the mattress The plug is fastened to the mattress by a strap If the pump is operating when the CPR is...

Страница 27: ...21 Make sure the plug is securely fitted 1 2...

Страница 28: ...n sequence the patient body is replaced by an egg timer icon 2 When the seat cushion inflation is complete the default seat operating screen is displayed the egg timer is replaced by the patient body...

Страница 29: ...rom the mains remove the plug from the mains power socket 2 Disconnect the tubeset from the pump refer to Disconnecting the Tubeset on page 10 3 Depress the two valves on the underside of the seat cus...

Страница 30: ...hlorine this may vary from 250ppm to 10 000ppm depending on local policy and contamination status Wipe all cleaned surfaces with the solution then wipe with a cloth moistened in water and dry thorough...

Страница 31: ...the mattress top cover including laundering guidelines refer to COVER SPECIFICATION on page 30 Re use with multiple patients Professional hygiene maintenance is required before re use of the system wi...

Страница 32: ...d to abnormal treatment e g immersed in water or dropped the unit must be returned to an authorised service centre Biofilter The internal biofilter can be run continuously for 12 months before it requ...

Страница 33: ...is securely pushed onto the pump until a click is heard Low Pressure 1 The tubeset is not connected properly 2 CPR control not fully closed 3 There is a leak in the system 1 Remove and reconnect tubes...

Страница 34: ...PUMP Symbols The operator must read this document Instructions for Use before use Note This symbol is blue on the product label Run Standby Button Do not dispose of in domestic refuse With respect to...

Страница 35: ...SIZE INFORMATION Part No Description Width 465006ADV Alpha RESPONSE Mattress Replacement Glide IS2 USA 34 9in 886mm 465006DAR Alpha RESPONSE Mattress Replacement Reliant IS2 USA 34 9in 886mm 465008ADV...

Страница 36: ...40 F or air dry Tumble dry only at 60 C 140 F Max Drying Temperatures Max 80 C 176 F Max 80 C 176 F Wipedown Chemicals b Chlorine at strength of 1000ppm or Alcohol at 70 concentration no phenol ensure...

Страница 37: ...nt in which radiated RF disturbances are controlled The customer or the user of the Alpha Response pump can help prevent electromagnetic interference by maintaining a minimum distance between portable...

Страница 38: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Страница 39: ...2 745 866 E mail international arjohuntleigh com ITALIA ArjoHuntleigh S p A Via Giacomo Peroni 400 402 IT 00131 ROMA Tel 39 0 6 87426211 Fax 39 0 6 87426222 E mail Italy promo arjohuntleigh com NEDERL...

Страница 40: ...solutions for infection control within healthcare and contamination prevention within life sciences Maquet specializes in solutions therapies and products for surgical interventions interventional car...

Отзывы: