Page | 14 

 

Commercial in confidence. Information for Use Manual Revision COU-E-REV12-080920 

5.  PRODUCT CARE -

 Continued 

 

5.4 Troubleshooting 

 

Symptom 

Possible Cause 

Action 

All actuators fail to work 

Power disconnected. 

Ensure mains supply lead is 
connected 

Control box and Footswitch 
plugs disconnected 

Check the plugs are fully 
inserted in the control box. 

Blown fuse 

Contact your local Arjo 
Service Department. 

Damaged footswitch cable 

Replace footswitch 

Control box overheated 
due to excessive use 
exceeding duty cycle. 

Allow control box to cool 
down. 

 

One actuator fails to work

 

 

Actuator plug disconnected

 

Check that actuator plug is 

fully inserted in control box. 

 

Brakes or steering lock not 
effective

 

 

Wear or damage

 

 

Adjust or replace castor

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание Prioma Couch Electric

Страница 1: ...INSTRUCTIONS FOR USE COU E REV12 080920 Electric Prioma Couch Range...

Страница 2: ...t 8 4 3 Backrest Adjustment 8 4 4 Knee Positioning Adjustment 8 4 5 Battery Backup if fitted 8 4 6 Central Locking Castor System 9 4 7 Fold down Arms if fitted 10 4 8 Paper Roll Holder if fitted 11 5...

Страница 3: ...in the presence of flammable gases Do not use electrically operated couches in operating theatres Use the couch to support one patient only The Prioma couch is intended for one person use and may beco...

Страница 4: ...hospital lines and power cables are free from possible entanglement with mechanical parts of the couch Always ensure that attendant and patient limbs are free from potential entrapment areas before op...

Страница 5: ...tended Area of Use Prioma couches are appropriate for use in private practices nursing homes physio centres hospitals and clinics for the purpose of a wide range of examination and treatment procedure...

Страница 6: ...ks not shown in image D 125mm Central Locking Castors x 4 E ABS Plastic Crank Arm Covers F Handset Storage Bar both sides of couch G Central Lock Pedal Neutral Brake Tracking H Upholstered Foot Sectio...

Страница 7: ...e couch Disconnect the power supply cord from the electricity supply and store safely before moving the couch 3 1 Electricity Supply Connect the mains cable plug to a suitable socket outlet Make sure...

Страница 8: ...tween the outer extremities of the couch and other building fixtures and equipment when operating couch functions Adjustment of the couch occurs only while the button is pressed Adjustment will stop w...

Страница 9: ...t use Bed operation will be slower when the power cord is unplugged and couch is powered by the batteries CAUTION If battery backup is fitted and to ensure that the battery is kept fully charged and p...

Страница 10: ...d in any direction With the pedals in the Red Down Brake position the brakes are applied to all four castors of the couch simultaneously and the couch is immobilised With the pedals in the Green Down...

Страница 11: ...rioma 1 Section Arm Length 1000mm O A Fig 4 To Lower Grip and hold the Arm in the middle of the top rail A and pull up as indicated by the directional arrow fig 1 to unlock Controlling the Arm allow i...

Страница 12: ...processes on the Prioma Couch The following procedure also applies to the accessories 5 1 Cleaning Clean all surfaces with a disposable cloth soaked in a neutral detergent and hand hot water Start by...

Страница 13: ...the power supply plug is not damaged If either the power cable or plug is damaged then both the cable and plug must be replaced as a complete assembly by an approved service agent CASTORS 1 Apply the...

Страница 14: ...s should take no more than one 1 minute to complete Visual Inspection of Control Panel Handset Casing 1 1 Inspect the outer plastic casing for any obvious cracks or damage 1 2 Inspect for damage to th...

Страница 15: ...otswitch plugs disconnected Check the plugs are fully inserted in the control box Blown fuse Contact your local Arjo Service Department Damaged footswitch cable Replace footswitch Control box overheat...

Страница 16: ...0 Kg 8 0 Kg With Fold down Arms Paper Roll Holder fitted excluding paper roll 13 5 Kg 9 5 Kg 9 5 Kg Safe Working Load SWL 200 Kg 200 Kg 200 Kg Maximum Patient Weight 165 Kg 165 Kg 165 Kg Electrical D...

Страница 17: ...ation Label Safe Working Load Refer to Instructions for Use Date of Manufacture Product Reference Number 11 Digit Code Representing Various Options Build Types Product Serial Number Manufacturer Indic...

Страница 18: ...ularly gas springs actuators batteries and other electrical devices may produce substances that are hazardous to the environment To minimise these hazards contact Arjo for information on correct dispo...

Страница 19: ...regarding this product please contact your local Arjo office or approved distributor Arjo offices can be found at www arjo com int region select Have the model number and serial number of the equipme...

Страница 20: ...ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to...

Отзывы: