25
'LVSOD\GXULQJRSHUDWLRQOLQH
The line is used to set what is displayed on the display during
operation.
• 1
=
Temperature
• 2
=
A
0
-value
• 3
=
A
0
-value / Temperature
• 4 = Remaining time
• 5 = Temperature / Remaining time
• 6
=
A
0
-value / Remaining time
• 7
=
A
0
-value / Remaining time / Temperature
• Default value is 7
Changing the settings requires authorization level C.
4.11 Program
statistics
0D[GLVLQIHFWLRQWHPSHUDWXUHOLQH
The line shows the highest temperature reached (5–100°C) in the
last program that was run. This value cannot be changed.
1XPEHURISURJUDPVVWDUWHGOLQH
The line shows the total number of programs started (00000–99999)
since the machine was installed. This value cannot be changed.
3URJUDPWLPHWRWDOOLQH
The line shows the total duration (00000–99999 hours) of all
programs run since the machine was installed. This value cannot
be changed.
4.12 Function
test
5HDGLQJLQSXWVOLQHV±
The lines show the status of the digital inputs. Values cannot be
changed.
Line
Input
Function
51
01
Door closed, 0 = open, 1 = closed
52
02
Door locked, 0 = unlocked, 1 = locked
53
03
Leakage sensor, 0 = OK, 1 = leak
54
04
Rinse-aid low, 0 = OK, 1 = low (open)
55
05
Descaler low, 0 = OK, 1 = low (open)
56
06
Level, C tank, 0 = no, 1 = yes (open)
57
07
Process agent low, 0 = OK, 1 = low
(open)