18
Line
Parameter
6HWWLQJ
basic
6HWWLQJ
range
$XWKRUL]DWLRQ
Lines 15-16: Disinfection
15
Disinfection temperature
91
80–95°C (Note: Celsius only) C
16
Disinfection time
60
0-1,800 seconds
C
17
Preheating
50
0-99 0 = switched off
C
18
Run time for ventilator after
process
120
0-600 seconds
C
19
Inactivity warning
1
0 = Inactive 1 = Active
C
Lines 21-33: Error statistics
21
Counter for error code F1
0-99 errors
C
24
Counter for error code F4
0-99 errors
C
27
Counter for error code F7
0-99 errors
C
28
Counter for error code F8
0-99 errors
C
29
Counter for error code F9
0-99 errors
C
30
Counter for error code F10
0-99 errors
C
31
Counter for error code F11
0-99 errors
C
33
Counter for error code F13
0-99 errors
C
Line 37: Display
37
Display during operation
7
1 = Temp, 2 = A
0
,
3 = Temp + A
0
,
4 = Remaining time,
5 = Time+Temp, 6 = A
0
+time,
7 = A
0
+Temp+Time
C
Lines 40-42: Program statistics
40
Max. temperature
5–100°C (displayed with
chosen temp scale)
41
Counter for number of
processes
0-99999
42
Counter, operating time
0-99999 h
Lines 51-76: Function test
51
Input 1 - Door closed
0 = open, 1 = closed
-
52
Input 2 - Door locked
0 = unlocked, 1 = locked
-
53
Input 3 - Leakage sensor
0 = OK, 1 = leakage
-
54
Input 4 - Rinse-aid low
0 = OK, 1 = low
-
55
Input 5 - Descaler low
0 = OK, 1 = low
-
56
Input 6 - Level, C tank
0 = no, 1 = yes
-
57
Input 7 - Process agent low
0 = OK, 1 = low
-
61
Output 1 - Cold water
0 = off, 1 = on
BC