38
Care and Maintenance
Inspection for lift
cassette
Initially
Before
every use
Every two
months or
500
cycles
Every
four
months
or 1000
cycles
Every
year or
2500
cycles
Every two
years or
5000
cycles
Inspect connecting joints for
proper attachment (trolley and
spreader bar).
X
Verify that the emergency brake
on the drum is turning freely.
X
Verify the emergency brake.
X
Verify emergency lowering
mechanism.
X
Verify alternative up and down
buttons on cassette.
X
Load test with the SWL (safe
working load) recommended.
X
FREQUENCY
Inspections for rails
Initially
Before
every use
Every two
months or
500
cycles
Every
four
months
or 1000
cycles
Every
year or
2500
cycles
Every two
years or
5000
cycles
Torque end stoppers to 20 N. m.
(15 lbf.ft).
X
X
Make sure that the bracket
locking device is not visible.
X
X
Make sure rail joints are closed
and that the spring pins are
centered.
X
X
Make sure the rail is straight
when it is not loaded.
X
X
Make sure the adjusted load
setting of the lift is equal or
lower than the safe working
load of the installation.
X
X
Check that the accessories
(turntable and exchanger) are
complete and correctly
maintained.
X
X
Make sure that the attachments
(ceiling brackets, wall post,
wall brackets) have not been
displaced.
X
X
Inspect track end stoppers.
Inspect and tighten hardware (if
necessary).
X
Load test with the SWL (safe
working load) recommended.
X
FREQUENCY
NOTE: If the product does not work as intended, immediately contact your local Arjo
representative for support.
Содержание MAXI SKY 600
Страница 23: ...23 How to Use the Maxi Sky600 Ceiling lift Fig 11 Attachment clips Step 1 Step 2 Step 3 Step 4 Step 5 Step 6...
Страница 25: ...25 How to Use the Maxi Sky600 Ceiling lift Fig 12 Attachment clips Step 1 Step 2 Step 3 Step 4 Step 5...
Страница 31: ...31 How to Use the Maxi Sky600 Ceiling lift Fig 22 Fig 23 Lying position Touch pad Sitting position...
Страница 52: ......
Страница 53: ......