39
Care and Maintenance
Daily Checklist
The following procedures must be followed before
each use:
•
Charge the batteries. Park the lift on the charger
station whenever the lift is not in use.
•
Inspect the lift for any damage. If the lift casing
does not look properly aligned, or there are any
cracks or other damage on the lift, or there are
parts missing, do not use it. Contact your Arjo
representative to have the lift serviced.
•
Inspect the strap for any visible signs of wear,
frays, loose threads or other damage. If there is
any evidence of damage, do not use it. Contact
your local Arjo representative to have the lift ser-
viced.
•
Inspect the sling for tears, frayed straps or loose
stitching. If the sling has any of the above dam-
age, do not use it. Contact your local Arjo repre-
sentative to have the sling replaced.
•
Inspect the spreader bar for any signs of cracking
or damage.
•
Ensure the split ring and cotter pins that attach the
spreader bar to the strap are secured.
Inspection and Cleaning
To clean the
Maxi Sky
600
, wipe it down with a damp
cloth using warm water and a disinfectant cleaner.
Disinfectant wipes, supplied already impregnated
with a 70% v/v solution of isopropyl alcohol, can
also be used.
Rub the lift vigorously when using the wipes, to
promote an effective disinfection of its entire surface.
Do not use phenol, chlorine or any other type of
solvent that may damage the finish.
To ensure a better rolling surface for the trolley
wheels, clean the inside of the track every 4 months.
To do so, insert a damp cloth in the opening and slide
it from one end of the track to the other.
WARNING: Before each use, make sure all
end stoppers are in place and secured.
Strap Inspection
If the strap is damaged or shows signs of wear or
discoloration, the acceptable load on the strap before
rupture can drop rapidly and present a danger for the
patient or caregiver. Arjo recommends a thorough
inspection of the straps every 2 months as follows:
1) Completely unwind the strap.
2) Look for any signs of wear or discoloration (see
Fig. 31 below).
Fig. 31
WARNING: Always reinstall the rail end
stoppers (if removed) after servicing.
Loose threads in stitched
areas
Noticeable discoloration
(strap color is lighter
than color in the stitched
area)
Edge wear (fraying)
Middle wear
WARNING: If there is any sign of wear as
indicated above or any other visual
defects, the strap must be inspected by
qualified personnel and changed if
required. By continuing to use the lift
without changing a damaged strap, the
safety of the caregiver or patient is greatly
compromised.
NOTE: The manufacturer recommends
changing the strap at least every two years
(see “Care and Maintenance” section). By
continuing to use the lift without changing
the strap, the safety of the caregiver or
patient is greatly compromised.
Содержание MAXI SKY 600
Страница 23: ...23 How to Use the Maxi Sky600 Ceiling lift Fig 11 Attachment clips Step 1 Step 2 Step 3 Step 4 Step 5 Step 6...
Страница 25: ...25 How to Use the Maxi Sky600 Ceiling lift Fig 12 Attachment clips Step 1 Step 2 Step 3 Step 4 Step 5...
Страница 31: ...31 How to Use the Maxi Sky600 Ceiling lift Fig 22 Fig 23 Lying position Touch pad Sitting position...
Страница 52: ......