21
EN
12. Speci
fi
cations
12.1 Equipment
Classi
fi
cation
Type of protection against
electric shock
Class II, Double Insulated
Degree of protection against
electric shock
Type BF
Mode of operation
Continuous
Degree of protection against
solid and liquid ingress
IP21 - Protection against ingress of solid objects more
than 12.5mm diameter and water droplets falling
vertically.
Degree of safety of
application in the presence of
a
fl
ammable anaesthetic
Equipment not suitable for use in the presence of
fl
ammable gases or oxygen rich environments.
12.2 General
Model
WoundExpress
™
Therapy Device
Part Numbers
WE100P
Pressure Range
60 mmHg ± 5mmHg
Supply voltage
230 V AC
Supply Frequency
50Hz
Pump Fuse Rating
F500 mA 250 V
Power input
14 VA
Case Material
Fire Retardant ABS Plastic
Size
270 x 130 x 150 mm (10.6 x 5.1 x 5.9”)
Weight
2.5 kg (5.5 lb)