2. Introduction

The WoundExpress Therapy Device is a system consisting of a pump and garment 
which operates by using low pressure air to deliver a compression therapy.

You must read and fully understand this manual before using the system.

Use this manual to initially set up the system, and keep it as a reference for day-to-day 
routines and as a guide to maintenance.

If you have any dif

fi

 culties in setting-up or using the WoundExpress Therapy Device, 

contact your local Huntleigh representative.

2.1 Intended 

Use

The intended use of this product is to manage the list of clinical conditions detailed in 
the “Indications” (section 3.1).
The WoundExpress Therapy Device should be used as part of a prescribed plan of 
care detailed in the “Indications” section.

2.2 

About the WoundExpress Therapy Device

The pump operates on a 4 minute automatically timed cycle, consisting of a 2 minute 
venous emptying phase and a 2 minute rest phase. The venous emptying phase 
consists of six 20 second compression cycles, while no compression takes place during 
the rest phase. Each 20 second compression cycle is characterised by overlapping 
synchronous in

fl

 ations and de

fl

 ations of the distal (lower), central (middle)and proximal 

(upper) chamber.

Compression Cycle

Delay to cycle start

In

fl

 ated state

De

fl

 ated state

Chamber 1 (lower)

0s

15s

5s

Chamber 2 (middle)

5s

10s

10s

Chamber 3 (upper)

12s

10s

10s

The maximum pressure in each chamber is 60mmHg. 

The segments within the garments are designed to prevent ridging and ensure high 
patient comfort and concordance.

A full technical description of the WoundExpress Therapy Device can be found in the 
Service Manual, part No. 785345, available from your local Huntleigh Service dept.

2.3 Use 

Environment

WoundExpress Therapy Device is suitable for use in hospital, primary care, community 
settings and in the home. It must not be used outdoors, or in any environment where it 
may come into contact with water.

Содержание Huntleigh WoundExpress

Страница 1: ...WoundExpress Therapy Device INSTRUCTIONS FOR USE WoundExpress WoundExpress 785344EN 2...

Страница 2: ...traindications 7 4 Preliminary Checks 8 5 Clinical Treatment Guide 9 6 Pump Description 10 6 1 Status Indicator Description 11 7 Operation 12 7 1 Garment Description 12 7 2 Applying the Garment 12 7 3...

Страница 3: ...18 9 2 1 General Care Maintenance and Inspection 18 9 2 2 Serial Labels 18 10 Trouble Shooting 19 11 Accessories 20 12 Specifications 21 12 1 Equipment Classification 21 12 2 General 21 12 3 Environme...

Страница 4: ...s if used in the presence of flammable gases Do not operate the unit from the mains supply if the mains cable is damaged Do not immerse any portion of the unit in water or other liquids Use only recom...

Страница 5: ...ct the pump from the mains power socket before cleaning and inspecting Make sure the system is clean and dry prior to use or storage Do not expose the system to naked flames such as cigarettes etc Do...

Страница 6: ...sisting of a 2 minute venous emptying phase and a 2 minute rest phase The venous emptying phase consists of six 20 second compression cycles while no compression takes place during the rest phase Each...

Страница 7: ...leg ulcers and mixed aetiology leg ulcers IPC may also be beneficial in the management of Lower limb pain Selection should be based upon a holistic assessment of the patients individual care needs No...

Страница 8: ...uncertain whether you have any of the above conditions please consult a physician before use CAUTION IPC should be used with care in patients with the following symptoms or conditions Peripheral neuro...

Страница 9: ...the unit is received Should any damage be evident or any parts missing ensure that Huntleigh is informed at once Storage If the unit not be required for immediate use it should be re sealed into its o...

Страница 10: ...herapy Device has a preset pressure of 60mmHg The recommended therapy time is 2 continuous hours per day A diary is included with the pump for the patient to complete to monitor concordance with the t...

Страница 11: ...or Indicates the pressure in the garment 3 Tube connectors For garment attachment 3A and 3B 4 Carry Handle For easy handling of the pump 5 Power status LED Indicates the power status of the pump Note...

Страница 12: ...compression cycle This can be during the 2 minute rest phase or when the device is turned off 2 The green zone indicates the proper functioning of the device The needle will fluctuate within this are...

Страница 13: ...tart date can be written onto the garment in the area provided 1 shown in diagram below using a ball point pen PATIENT NAME THERAPY START DATE __ __ __ __ __ __ 1 Huntleigh Healthcare Ltd 35 Portmanmo...

Страница 14: ...TIENT NAME THERAPY START DATE __ __ __ __ __ __ Huntleigh Healthcare Ltd 35 Portmanmoor Road Cardiff CF24 5HN UK TEL 44 0 2920 485885 www huntleigh diagnostics com 0088 LATEX SINGLE PATIENT USE WoundE...

Страница 15: ...the pump is in operation as you may damage the garment WARNING Patients must not walk or stand when wearing leg garments 7 3 Switching On The pump should be placed securely on a flat surface Connect t...

Страница 16: ...by 2min of deflation The recommended therapy time is 2 continuous hours per day A diary is included with the pump for the patient to complete to monitor concordance with the therapy It is normal durin...

Страница 17: ...op the patient therapy Note If it is required to completely isolate the pump from the mains power remove the plug from the mains power socket 7 6 Removing the Garment Make sure the pump power switch 1...

Страница 18: ...ing the cleaning process Do not spray cleaning solutions directly onto the pump Do not immerse the tubeset in water 8 1 Cleaning Clean all exposed surfaces and remove any organic debris by wiping with...

Страница 19: ...s Therapy Device pump is serviced every 24 months by an Huntleigh authorised service agent 9 2 WoundExpress Therapy Device Pump 9 2 1 General Care Maintenance and Inspection Check all electrical conne...

Страница 20: ...r Pump operates but garment will not inflate Blockage in garment supply tube Garment not fitted correctly to pump Air leak in garment Ensure that the tube airway is clear Check for kinked tubing Check...

Страница 21: ...20 EN 11 Accessories WARNING Use only recommended accessories listed in this manual Accessory Order Code WoundExpress Therapy Device Garment WE100G...

Страница 22: ...12 5mm diameter and water droplets falling vertically Degree of safety of application in the presence of a flammable anaesthetic Equipment not suitable for use in the presence of flammable gases or ox...

Страница 23: ...hPa Storage and transport short term 25 C to 70 C 13 F to 158 F 20 to 95 500 to 1060 hPa Note When exposed to extreme temperature during storage the pump must be allowed to adjust to normal temperatur...

Страница 24: ...al procedures Refer to this document Instructions for Use for a description of the product classification General Warning This symbol signifies that this product complies with the essential requiremen...

Страница 25: ...quipment into a different outlet so that the devices are on different branch circuits WARNING If this equipment needs to be used adjacent to other electrical equipment normal operation must be checked...

Страница 26: ...mined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of the equipment marked with the following symbol NO...

Страница 27: ...tput power of transmitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 0 35 P 800MHz to 2 5GHz d 0 7 P 0 01 0 12 0 04 0 07 0 1 0 38 0 11 0 22...

Страница 28: ...61000 4 5 1 kV line s to line s 1 kV line s to line s Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on pow...

Страница 29: ...his symbol signifies that this product including its accessories and consumables is subject to the WEEE Waste Electrical and Electronic Equipment regulations and should be disposed of responsibly in a...

Страница 30: ...n suitable packing Attach a decontamination certificate or other statement declaring that the product has been cleaned to the outside of the package Mark the package Service Department For further det...

Страница 31: ......

Страница 32: ...ous improvement we reserve the right to modify designs without prior notice and are trademarks of Huntleigh Technology Limited A Member of the Arjo Family AW 1001008 4 As a proud member of the Arjo fa...

Отзывы: