14. Electromagnetic
Compatibility
Make sure the environment in which WoundExpress Therapy Device is installed is
not subject to strong sources of electromagnetic interference (e.g. radio transmitters,
mobile phones).
This equipment generates and uses radio frequency energy. If not installed and used
properly, in strict accordance with the manufacturer’s instructions, it may cause or
be subject to interference. Type-tested in a fully con
fi
gured system, complies with
EN60601-1-2, the standard intended to provide reasonable protection against such
interference. Whether the equipment causes interference may be determined by
turning the equipment off and on. If it does cause or is affected by interference, one or
more of the following measures may correct the interference:
•
Reorienting the equipment
•
Relocating the equipment with respect to the source of interference
•
Moving the equipment away from the device with which it is
interfering
•
Plugging the equipment into a different outlet so that the devices are
on different branch circuits
WARNING
If this equipment needs to be used adjacent to other electrical
equipment, normal operation must be checked before use.
Guidance and Manufacturer’s declaration - electromagnetic emissions
The WoundExpress Therapy Device is intended for use in the electromagnetic environment speci
fi
ed below.
The customer or the user of the WoundExpress Therapy Device should assure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environment - guidance
RF emissions
CISPR 11
Group 1
The WoundExpress Therapy Device uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The WoundExpress Therapy Device is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fl
uctuations /
fl
icker emissions
IEC 61000-3-3
Complies