50
TROUBLESHOOTING
50
TROUBLESHOOTING
Troubleshooting
PROBLEM DESCRIPTION
ACTION
Nothing happens when pushing a
button.
Check that the emergency stop is not pushed down. Turn
emergency stop clockwise until it pops up to reset.
The battery level LED is turned on but
the main LED is turned off.
Power failure. Contact technician.
1RZDWHUÀRZRUEDGZDWHUÀRZGXULQJ
&ODVVLF/LQHEDWKWXE¿OOLQJDQG
shower.
1.
&KHFNLQFRPLQJZDWHUÀRZ
2.
&KHFNFOHDQ¿OWHUV
1RZDWHUÀRZRUEDGVSUD\SDWWHUQ
during shower only.
1RZDWHUÀRZ
• Turn the shower off if no water comes out from the shower
KDQGOHZLWKLQVHFRQGVFRQWDFWTXDOL¿HGSHUVRQQHO
Bad spray pattern:
• Check shower handle. Replace if necessary.
The water temperature can not be
regulated.
Check temperature on incoming water. Check that the water
temperature settings knob is working as intended.
Scalding protection is not working when
temperature display shows 3 lines and
not the temperature:
--.-°C
&RQWDFWTXDOL¿HGSHUVRQQHO
NOTE
Before starting the troubleshooting check: power supply, water supply and the LED indicators on
the control panel. If the problem can not be solved with the problem solving actions below, please
contact an Arjo authorised technician.