44
CLEANING AND DISINFECTION
Disinfect (step 25 to 39 of 39)
25. Press the Disinfection ON/OFF button for 5
seconds.
For bathtubs with Air Spa system:
Approximate time for the automatic disinfection
is 3 to 4 minutes. 10 seconds after the start, the
disinfection spray handle is activate for 2 minutes
before it shuts off.
26. Spray disinfectant into the drain.
27. Close the Drainage plug.
28. Spray disinfectant on all internal bath surfaces,
and at the footrest at the bottom of the tub. Turn
over the footrest and spray on the other side. Pay
attention to parts, knobs and sensors attached to
the bath surface such as drain control knob and
grab handles. See Figure 74.
29. Soak a cloth in disinfectant.
30. Wipe all other contact areas outside of the bath
(particularly behind the head), shower handle and
Water temperature setting knob with the cloth.
31. Allow
an appropriate contact time
of the
disinfectant according to the disinfectant IFU.
32. For bathtubs with Air Spa system:
Disinfection cycle will turn off automatically.
For bathtubs without Air Spa system:
Press the Disinfection ON/OFF button on the
control panel to end the disinfection process.
33. Put the Disinfection spray handle back and put
the covers back on.
34. Open the Drainage plug.
35. Rinse all internal parts clean with water having a
temperature of approximately 25 °C (77 °F) using
the shower handle to rinse off the disinfectant.
See Figure 75.
Figure 75
Figure 74
Содержание Classic Line 2
Страница 1: ...INSTRUCTIONS FOR USE Classic Line 2 04 ACL 00_1EN 08 2020 ...
Страница 6: ...Intentionally left blank ...